Accurate and Rapid Measurement of Glycated Hemoglobin Using HLC-723 G11 Variant Mode

机译:使用HLC-723 G11变异模式准确快速地测量糖化血红蛋白

获取原文
           

摘要

Background: Type II diabetes mellitus causes many complications, and its prevalence continues to increase in Korea. Accurate measurement of glycated Hb (HbA1c) is important because of its usefulness in diagnosis, follow-up, and prediction of prognosis. We tested the analytical performance of the HLC-723 G11 Variant Mode (G11vr; Tosoh Bioscience, Inc., Tokyo, Japan), recently introduced to Korea, in detecting HbA1c. Methods: We evaluated precision, linearity, carry-over, and turnaround time. Using 208 samples, including 108 flagged samples, we compared HbA1c concentrations from four analyzers through correlation analysis: G11vr, HLC-723 G8 Variant Mode (G8vr, Tosoh Bioscience), HLC-723 G11 Standard Mode (G11st, Tosoh Bioscience), and HLC-723 G8 Standard Mode (G8st, Tosoh Bioscience). We used HPLC mass spectrometry (MS) and capillary electrophoresis (CE) to confirm the HbA1c concentrations of 15 additional known Hb variant samples. Results: Repeatability (% CV) in measuring low- and high-concentration controls was 0.57% and 0.35%, respectively; within-laboratory precision was 0.86% and 0.69%, respectively. In a linearity test, the coefficient of determination was 0.9999 (measurement range: 3.64% to 18.59%) for HbA1c. The correlations between G11vr and other analyzers were weaker for flagged samples than for non-flagged samples. The carry-over effect was less than 0.4%. Turnaround time for a single sample was lower in G11vr (one minute) than in G8vr (1.6 minutes). For 15 samples with Hb variants, G11vr HbA1c results were more similar than those of other analyzers to HPLC-MS and CE results. Conclusions: G11vr showed adequate performance and rapid turnaround time in measuring HbA1c.
机译:背景:II型糖尿病引起许多并发症,在韩国其患病率持续上升。糖化血红蛋白(HbA1c)的准确测量非常重要,因为它可用于诊断,随访和预测预后。我们测试了最近引入韩国的HLC-723 G11变异模式(G11vr; Tosoh Bioscience,Inc.,日本东京)在检测HbA1c中的分析性能。方法:我们评估了精度,线性,残留和周转时间。我们使用了208个样本(包括108个标记样本),通过相关分析比较了四个分析仪的HbA1c浓度:G11vr,HLC-723 G8变异模式(G8vr,Tosoh Bioscience),HLC-723 G11标准模式(G11st,Tosoh Bioscience)和HLC -723 G8标准模式(Tosoh Bioscience,G8st)。我们使用HPLC质谱(MS)和毛细管电泳(CE)来确认15个其他已知Hb变异样品的HbA1c浓度。结果:低浓度和高浓度对照的重复性(%CV)分别为0.57%和0.35%。实验室内的精确度分别为0.86%和0.69%。在线性测试中,HbA1c的测定系数为0.9999(测量范围:3.64%至18.59%)。对于标记的样本,G11vr与其他分析仪之间的相关性要弱于未标记的样本。残留效应小于0.4%。在G11vr中,单个样品的处理时间(一分钟)比在G8vr中的处理时间(1.6分钟)要短。对于15个具有Hb变体的样品,G11vr HbA1c结果与其他分析仪的HPLC-MS和CE结果更相似。结论:G11vr在测量HbA1c中表现出足够的性能和快速的周转时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号