We aim to maximize the pumping volume of a pulsatile ventricular assist device, where the diaphragm is covered with an endothelial cell layer. These cells are estimated to survive a cyclic strain up to fifteen percent. To increase the pumping volume under this strain constraint we use an approach based on corrugation of the diaphragm in its reference configuration. The paper explains the parametrization scheme for finding corrugation shapes, addresses modeling and evaluation schemes and reports on the results of a parameter study. The results show that corrugated diaphragm shapes are effective for increasing pumping volumes under a strain constraint.
机译:我们的目标是最大程度地增加脉动心室辅助设备的泵送容积,在该设备中,隔膜被内皮细胞层覆盖。估计这些细胞能在高达15%的循环应变下存活。为了在这种应变约束下增加泵送量,我们使用了一种基于膜片在其参考配置中起皱的方法。本文介绍了用于查找波纹形状的参数化方案,介绍了建模和评估方案,并报告了参数研究的结果。结果表明,在应变约束下,波纹膜片形状对于增加泵送量是有效的。