首页> 美国卫生研究院文献>other >A Thrombus Susceptibility Comparison of Two Pulsatile Penn State 50cc Left Ventricular Assist Device Designs
【2h】

A Thrombus Susceptibility Comparison of Two Pulsatile Penn State 50cc Left Ventricular Assist Device Designs

机译:两种脉动Penn状态50CC左心室辅助装置设计的血栓敏感性比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Left ventricular assist devices (LVADs) have proven successful as bridge to transplant devices for patients awaiting donor organs. While survival rates continue to increase, destination therapy remains hindered by thrombus formation within the device. Research has shown that thrombosis is correlated to the fluid dynamics within the device and may be a result of sustained shear rates below 500 s−1 on the polyurethane blood sac used in the Penn State pulsatile LVAD. Particle image velocimetry is used to compare flow within two 50 cc LVAD designs to assess fluid patterns and quantify wall shear rates in regions known from in vivo studies to be susceptible to thrombus formation. The two designs differ in their front face geometry. The V-1 model has an outward facing “dome” whereas the face of the V-2 model is flat. A thrombus susceptibility metric, which uses measured wall shear rates and exposure times, was applied to objectively compare pump designs over the entire cardiac cycle. For each design, there are regions where wall shear rates remained below 500 s−1 for the entire cardiac cycle resulting in high thrombus susceptibility potential. Results of this study indicate that the V-2 device had an overall lower propensity for thrombus formation in the current region of interest.
机译:左心室辅助设备(LVAD)已被证明可以成功地成为等待供体器官的患者移植设备的桥梁。尽管存活率继续提高,但目的地治疗仍然受到装置内血栓形成的阻碍。研究表明,血栓形成与设备内的流体动力学有关,并且可能是宾夕法尼亚州搏动性LVAD所使用的聚氨酯血囊的持续剪切速率低于500 s -1 的结果。粒子图像测速仪用于比较两个50 cc LVAD设计中的流量,以评估流体模式并量化体内研究已知易形成血栓的区域中的壁剪切速率。两种设计的正面几何形状不同。 V-1模型具有面向外的“圆顶”,而V-2模型的表面是平坦的。使用血栓敏感性指标(使用测得的壁剪切速率和暴露时间)客观比较整个心动周期中的泵设计。对于每种设计,在整个心动周期中都有一些区域的壁剪切率保持在500 s -1 以下,从而导致较高的血栓易感性。这项研究的结果表明,V-2装置在当前感兴趣的区域总体上具有较低的血栓形成倾向。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号