Angio-OCT 与 FD-OCT 的眼底检查相关性分析

         

摘要

目的:比较两种光学相干断层成像术( optical coherence tomography ,OCT):血流成像光学相干断层成像术( Angio-optical coherence tomography , Angio-OCT)和傅立叶域光学相干断层成像术( Fourier domain-optical coherence tomography , FD-OCT)眼底检查的相关性及差异性。方法:分别用Angio-OCT和FD-OCT测量20例受试者双眼神经纤维层厚度、视盘各参数和节细胞复合体厚度等,并对两种仪器测量值进行比较,采用Pearson相关分析研究两种仪器测量值的相关性,用配对样本t检验比较两种仪器测量值的差异性。结果:(1)两种OCT所测RNFL厚度、视盘参数及GCC厚度,总体相关性为0.7~0.8。(2) Angio-OCT 所测GCC厚度比FD-OCT所测平均值高,但其它参数如RNFL厚度、视盘面积等,Angio-OCT的测量值普遍比FD-OCT的测量值低。结论:两种OCT的测量值( RNFL厚度、视盘参数、GCC厚度)有一定相关性,但不能直接比较。%?AIM:To compare the coherence and difference on the fundus examination made with two kinds of optical coherence tomography ( OCT): Angio-OCT and fourier domain-optical coherence tomography ( FD-OCT) .?METHODS:Using Angio-OCT and FD-OCT to measure the retinal nerve fiber layer ( RNFL ) thickness, optic parameters, and ganglion cell complexes ( GCC ) thickness from 20 subjects respectively.The coherence was tested with Pearson's correlation coefficient, the difference was tested with paired Student t testing.?RESULTS:The total correlation of the RNFL thickness, optic parameters, GCC thickness made with two kinds of OCT was between 0.7-0.8;the RNFL thickness, optic disk area etc.made with the Angio-OCT were lower than those made with FD-OCT except for the GCC thickness.?CONCLUSION: The results made with two kinds of OCT from the same subject has certain coherence, but cannot be compared directly.