Recently, a new PLZT bulk single plate, called ceramic-controlled piezoelectric with two Pt wires (CCP2) has been produced; this CCP2 has two (300 μm of diameter and 1 cm long) Pt-wire implants. These isolation was validated using four experimental setups, one of them determine that ceramic-controlled piezoelectric with two Pt wires (CCP2) support up to 6 kV DC before it cracks (short cut). The second experimental setup determined high resistance about 3.9×109Ω and 1.8×109Ω measured on lateral sides and among Pt wires of CCP2 respectively. The third experimental setup was to obtain the current leakage from CCP2 and it was 6 nA. The fourth experimental setup was to obtain frequency response that was the maximum up to 2.2 MHz and was a pass band filter.
机译:最近,已经产生了一种新的PLZT散装单板,称为具有两个Pt线(CCP2)的陶瓷控制的压电(CCP2); 该CCP2具有两个(300μ m型直径和1厘米长)pt-wire植入物。 使用四个实验设置验证了这些隔离,其中一个是用两个Pt线(CCP2)的陶瓷控制的压电在其裂缝之前支持高达6kV DC(短切)。 第二实验设置确定高电阻约3.9× 10 9 Ω 1.8× 10 9 Ω 在横向侧和CCP2的PT线上测量。 第三个实验设置是从CCP2获得电流泄漏,为6A。 第四个实验设置是获得最大高达2.2 MHz的频率响应,并且是通带过滤器。