Recently, a new PLZT bulk single plate, called ceramic-controlled piezoelectric with two Pt wires (CCP2) has been produced; this CCP2 has two (300 μm of diameter and 1 cm long) Pt-wire implants. These isolation was validated using four experimental setups, one of them determine that ceramic-controlled piezoelectric with two Pt wires (CCP2) support up to 6 kV DC before it cracks (short cut). The second experimental setup determined high resistance about 3.9×109Ω and 1.8×109Ω measured on lateral sides and among Pt wires of CCP2 respectively. The third experimental setup was to obtain the current leakage from CCP2 and it was 6 nA. The fourth experimental setup was to obtain frequency response that was the maximum up to 2.2 MHz and was a pass band filter.
机译:最近,已经生产出一种新的PLZT大块单板,称为带有两根Pt线(CCP2)的陶瓷控制压电板。该CCP2具有两个(直径300μm,长1 cm)的Pt线植入物。这些隔离已通过四种实验设置进行了验证,其中一种确定了带有两条Pt线(CCP2)的陶瓷控制压电在破裂前(短路)可支持高达6 kV的直流电。第二个实验装置确定在CCP2的Pt导线的侧面和侧面分别测得的高电阻分别约为3.9×10 9 Ω和1.8×10 9 Ω。第三个实验设置是从CCP2获得电流泄漏,电流泄漏为6 nA。第四个实验设置是获得高达2.2 MHz的最大频率响应并且是通带滤波器。