First Results from Position-Sensitive Quantum Calorimeters using Mo/Au Transition-Edge Sensors

机译:使用MO / AU过渡边缘传感器的位置敏感量子量热计的首先结果

获取原文

摘要

We report the first results from a high-energy-resolution imaging spectrometer called a Position-Sensitive Transition-Edge Sensor (PoST). A PoST is a quantum calorimeter consisting of two Transition Edge Sensors (TESs) on the ends of a long absorber to do one dimensional imaging spectroscopy. Comparing rise time and energy information, the position of the event in the PoST is determined. Energy is inferred from the sum of the two pulses. We have fabricated 7- and 15-pixel PoSTs using Mo-Au TESs and An absorbers. We have achieved 32 eV FWHM energy resolution at 1.5 keV with a 7-pixel PoST calorimeter.
机译:我们通过称为位置敏感的过渡边缘传感器(POST)的高能分辨率成像光谱仪报告第一结果。柱是量子量热计,其由长吸收器的末端组成的两个过渡边缘传感器(TESS),以进行一维成像光谱。比较上升时间和能量信息,确定了帖子中的事件的位置。从两个脉冲的总和推断出能量。我们使用Mo-Au TES和吸收器制造了7个和15像素柱。我们在1.5 kev下实现了32eP FWHM能量分辨率,带有7像素的瓶子量热计。