This paper presents a portable readout interface IC for electrochemical sensing. The circuit conducts the chronoamperometry measurement by generating two stimulus pulses. A current-to-frequency converter senses the current change from the bio-sensing transistor built by two gold electrodes at the gate. The results are measured by using a time-to-digital converter. The readout circuit consumes 14$mu$ W at a 1.2V supply, and its function is verified under the potassium ferricyanide test. The limit of detection is 1nM in a concentration range of 1mM.
机译:本文介绍了一种用于电化学传感的便携式读出接口IC。该电路通过产生两个刺激脉冲来进行计时测量测量。电流转换器感测到由​​栅极的两个金电极构建的生物传感晶体管的电流变化。通过使用时间到数字转换器来测量结果。读出电路在1.2V电源下消耗14美元 mu $ w,其功能在铁氰化钾试验下验证。检测极限在1mm的浓度范围内为1nm。