摘要

Magnetocardiographic (MCG) measurements in unshielded environment require efficient noise cancellation techniques. We have applied two software gradiometry methods to analyze the time series of signal and reference data recorded outside magnetic shielding with high temperature superconducting quantum interference device (HTS SQUID) based gradiometers. One method uses adaptive frequency dependent gradiometer coefficients determined in the Fourier domain to subtract the reference from the signal data. The other method combines recently developed techniques for nonlinear projection with properties of the wavelet transform to extract noise in state space. The analyzed MCG data sets showed improved signal-to-noise ratios for both methods as compared to the data recorded with the electronic gradiometer. In this way, it is possible to increase the bandwidth from 130 Hz for our electronic gradiometer to 250 Hz without using any additional filtering.
机译:非屏蔽环境中的心电图(MCG)测量需要有效的噪声消除技术。我们已经应用了两种软件梯度法,以基于高温超导量子干涉仪(HTS SQUID)的梯度计分析在磁屏蔽外部记录的信号和参考数据的时间序列。一种方法是使用在傅立叶域中确定的与频率相关的自适应梯度仪系数从信号数据中减去参考值。另一种方法将最近开发的非线性投影技术与小波变换的属性相结合,以提取状态空间中的噪声。与电子梯度仪记录的数据相比,两种方法的分析MCG数据集均显示出改进的信噪比。这样,可以将带宽从电子梯度仪的130 Hz增加到250 Hz,而无需使用任何其他滤波。