首页> 外文期刊>Perfusion >Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass.
【24h】

Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass.

机译:在体外循环过程中即时测量血细胞比容,血气,电解质,乳酸和葡萄糖的可靠性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Recently, the GEM Premier blood gas analyser was upgraded to the GEM Premier 3000. In addition to pH, pCO2, pO2, Na+, K+, Ca2+, and hematocrit measurement, glucose and lactate can be measured on the GEM Premier 3000. In this prospective clinical study, the analytical performance of the GEM Premier 3000 was compared with the Ciba Corning 865 analyser for blood gas/electrolytes/metabolites, and for hematocrit with the Sysmex XE 2100 instrument. METHODS: During a 6-month period, 127 blood samples were analysed on both the GEM Premier 3000 analyser and our laboratory analysers (Ciba Corning 865/Sysmex 2100 instrument), and compared using the agreement analysis for quantitative data. RESULTS: With the exception of K+, the other parameters (pCO2, pO2, Na+, Ca2+, hematocrit, glucose, and lactate) can be described in terms of the mean and standard deviation of the differences. For K+ measurement, a clear linear trend (r=0.79, p<0.001) in the deviation of the GEM Premier 3000 from the Ciba Corning was noticed, ie, in the lower or upper K+ reference range, the GEM Premier 3000 measured systematically too low or too high, respectively. Furthermore, in comparison with the other parameters, a therapeutically unacceptable systematic difference (mean of difference: -2.2%, p=0.05) in hematocrit measurement on the GEM Premier 3000 was observed for hematocrit values below 30%. The variance of the readings for the GEM Premier 3000 measurements was at clinically acceptable levels. CONCLUSION: The GEM Premier 3000 analyser seems to be suitable for point-of-care testing of electrolytes, metabolites, and blood gases during cardiopulmonary bypass. However, its downward bias in hematocrit values below 30% suggests that using the GEM Premier 3000 as a transfusion trigger leads to overtreatment with packed red cells.
机译:背景:最近,GEM Premier血气分析仪升级到GEM Premier3000。除了pH,pCO2,pO2,Na +,K +,Ca2 +和血细胞比容测量外,还可以在GEM Premier 3000上测量葡萄糖和乳酸。在这项前瞻性临床研究中,将GEM Premier 3000的分析性能与Ciba Corning 865分析仪的血气/电解质/代谢物以及血细胞比容进行了比较,并使用Sysmex XE 2100仪器进行了比较。方法:在6个月的时间里,在GEM Premier 3000分析仪和我们的实验室分析仪(Ciba Corning 865 / Sysmex 2100仪器)上分析了127个血液样本,并使用协议分析比较了定量数据。结果:除K +外,其他参数(pCO2,pO2,Na +,Ca2 +,血细胞比容,葡萄糖和乳酸盐)可以用差异的平均值和标准偏差来描述。对于K +测量,注意到GEM Premier 3000与Ciba Corning的偏差存在明显的线性趋势(r = 0.79,p <0.001),即,在K +的较低或较高参考范围内,GEM Premier 3000也进行了系统测量低或过高。此外,与其他参数相比,在血细胞比容低于30%的情况下,在GEM Premier 3000上观察到了在血细胞比容测定中治疗上无法接受的系统差异(均值:-2.2%,p = 0.05)。 GEM Premier 3000测量的读数差异在临床上可以接受。结论:GEM Premier 3000分析仪似乎适用于体外循环过程中电解质,代谢物和血气的即时检测。但是,其血细胞比容值的向下偏差低于30%,表明使用GEM Premier 3000作为输血触发会导致堆积红细胞过度治疗。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号