Comparison of point-of-care and central laboratory analyzers for blood gas and lactate measurements

机译:对血气和乳酸乳液测量的关怀和中央实验室分析仪的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Background Blood gas analysis and blood lactate measurement have important roles in patient management. Point-of-care (POC) testing simplifies and provides rapid blood gas and lactate measurements. This study aimed to compare pH, pCO(2), pO(2), and lactate measurements between a POC device and a benchtop blood gas analyzer typically used in a hospital central laboratory, and to evaluate the inter-device variability of the POC device. Methods A cross-sectional study was conducted with a sample size of 100. Each sample was measured for pH, pCO(2), pO(2), and lactate using a Nova pHOx plus L benchtop blood gas analyzer in the central laboratory and an i-STAT handheld POC device. The results of both devices were compared using Pearson or Spearman correlation coefficients and Bland-Altman tests. Testing of the inter-device variability was done by using three different i-STAT devices, and the results were compared statistically. Results Strong correlations were observed for all test results. In Bland-Altman analysis, >= 95% of the results were within the limits of agreement, with the exception of lactate, which had only 93%. The results that were beyond the limits were primarily lactate levels >8 mmol/L. Biases between the benchtop analyzer and the i-STAT were not clinically significant, except pH. No significant inter-device variability was observed between the i-STAT analyzers. Conclusion This comparison study of pH, pCO(2), pO(2), and lactate measurements between Nova pHOx plus L and i-STAT analyzers showed good agreement. However, lactate measurement results >8 mmol/L on the i-STAT analyzer should be interpreted with caution.
机译:背景技术血气分析和血液乳酸测量在患者管理中具有重要作用。护理点(POC)测试简化并提供了快速的血气和乳酸测量。本研究旨在比较POC设备和通常在医院中央实验室使用的台式血气分析仪之间的pH,PCO(2),PO(2)和乳酸测量,并评估POC设备的设备间变异性。方法采用100的样品大小进行横截面研究。使用中央实验室中的Nova Phox加上L L B台血液天然气分析仪测量每种样品的pH,PCO(2),PO(2),并在中央实验室中测量乳酸I-Stat手持POC设备。使用Pearson或Spearman相关系数和Bland-Altman测试进行比较两种器件的结果。通过使用三种不同的I-STAT设备进行设备间可变性的测试,结果在统计上进行比较。结果对所有测试结果观察到强烈相关性。在Bland-Altman分析中,> = 95%的结果是在协议的范围内,除乳酸外,只有93%。超出限制的结果主要是乳酸水平> 8mmol / L.除pH外,Benchtop分析仪和I-STAT之间的偏见并不临床显着性。 I-STAT分析仪之间未观察到显着的设备间可变性。结论Nova Phox Plus L和I-STAT分析仪之间pH,PCO(2),PO(2)和乳酸测量的这种比较研究表现出良好的一致性。但是,I-Stat分析仪上的乳酸乳酸测量结果> 8 mmol / L应该谨慎地解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号