Trapping 4-fluorobenzyl chloride in human plasma with chemical derivatization followed by quantitative bioanalysis using high-performance liquid chromatography/tandem mass spectrometry

机译:化学衍生法捕集人血浆中的4-氟苄基氯,然后使用高效液相色谱/串联质谱进行定量生物分析

获取原文
获取原文并翻译 | 示例
       

摘要

A quantitative bioanalytical method involving chemical derivatization, solid phase extraction (SPE) and high-performance liquid chromatography/tandem mass spectrometry (HPLC/MS/MS) was developed for the determination of 4-fluorobenzyl chloride (4FBCl) in human plasma. 4FBCl is a volatile and reactive molecule that is very unstable in human plasma. In order to stabilize 4FBCl in plasma samples prior to storage, 4-dimethylaminopyridine (DMAP) was added, forming a stable quaternary amine salt derivative. A three-step weak cation-exchange SPE procedure was then employed to remove excess DMAP. The plasma extracts were analyzed by HPLC/MS/MS using a TurboIonspray interface and multiple reaction monitoring. Unlike 4FBCl, the quaternary amine derivative shows excellent sensitivity in electrospray mass spectrometry. The method was validated over a concentration range of 0.5-500 ng/mL using 45 mu L of plasma. The maximum within-run and between-run precision observed in a three-run validation for quality control (QC) samples was 12.5 and 7.6%, respectively. The maximum percentage bias observed at all QC sample concentrations was 11.9%. The method has proven to be robust and compatible with high-throughput bioanalysis. Copyright (c) 2005 John Wiley A Sons, Ltd.
机译:建立了一种化学定量,固相萃取(SPE)和高效液相色谱/串联质谱(HPLC / MS / MS)的定量生物分析方法,用于测定人血浆中的4-氟苄基氯(4FBCl)。 4FBC1是挥发性和反应性分子,在人血浆中非常不稳定。为了在储存之前稳定血浆样品中的4FBCl,添加了4-二甲基氨基吡啶(DMAP),形成了稳定的季胺盐衍生物。然后采用三步弱阳离子交换SPE程序去除过量的DMAP。使用TurboIonspray接口和多反应监测,通过HPLC / MS / MS分析血浆提取物。与4FBCl不同,季胺衍生物在电喷雾质谱中显示出出色的灵敏度。使用45μL血浆在0.5-500 ng / mL的浓度范围内验证了该方法。在质量控制(QC)样品的三轮验证中观察到的最大轮内和轮间精度分别为12.5和7.6%。在所有QC样品浓度下观察到的最大百分比偏差为11.9%。该方法已被证明是可靠的,并且与高通量生物分析兼容。版权所有(c)2005 John Wiley A Sons,Ltd.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号