Artifacts at PET and PET/CT caused by metallic hip prosthetic material.

机译:金属髋关节假体材料引起的PET和PET / CT伪像。

获取原文
获取原文并翻译 | 示例

摘要

Hip prosthetic material and a steel rod were scanned in a water bath of fluorine 18 fluorodeoxyglucose (FDG) with positron emission tomographic (PET) and PET/computed tomographic (CT) scanners to evaluate the generation of artifacts adjacent to the metal. The influences of attenuation correction (AC), positioning of the object, and image reconstruction were examined. Use of CT- and germanium 68-based AC resulted in generation of artifacts that mimicked increased FDG uptake. These artifacts were more evident when the object was moved between the emission and transmission scans. When attenuation-weighted iterative reconstruction was used, these artifacts were less evident.
机译:髋假体材料和一根钢棒在氟18氟脱氧葡萄糖(FDG)的水浴中用正电子发射断层扫描(PET)和PET /计算机断层扫描(CT)扫描仪进行扫描,以评估与金属相邻的伪影的产生。研究了衰减校正(AC),物体放置和图像重建的影响。使用基于CT和锗68的交流电会产生伪影,这些伪影模仿了FDG的吸收增加。当物体在发射和透射扫描之间移动时,这些伪影更加明显。当使用衰减加权迭代重建时,这些伪像不太明显。