首页> 外文期刊>Medical and Biological Engineering and Computing: Journal of the International Federation for Medical and Biological Engineering >Event synchronous adaptive filter based atrial activity estimation in single-lead atrial fibrillation electrocardiograms
【24h】

Event synchronous adaptive filter based atrial activity estimation in single-lead atrial fibrillation electrocardiograms

机译:基于事件同步自适应滤波器的单导联心房颤动心电图心房活动估计

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

In this paper, an event synchronous adaptive filter (ESAF) is proposed to estimate atrial activity (AA) from a single-lead AF ECG in real time. The proposed ESAF is a kind of adaptive filter designed to have the reference fed with the impulse train synchronized with the R peak in a raw atrial fibrillation (AF) ECG and to input the timely delayed AF ECG into the primary input. To assess the performance, for ten simulated AF ECGs, the crosscorrelation coefficient (ρ) and the normalized mean square error (NMSE) between estimated AAs and ten original simulated AAs were calculated and, for ten real AF ECGs, the ventricular residue (VR) in QRS interval and similarity (S) in non-QRS interval were computed. As a result, these four parameters were revealed as ρ = 0.938 ± 0.016 and NMSE = 0.243 ± 0.051 for simulated AF ECGs and VR = 1.190 ± 0.476 and S = 0.967 ± 0.041 for real AF ECGs. These results were found to be better than those of the averaged beat subtraction (ABS) method, which had been previously considered the only way to estimate AA automatically in real time. In conclusion, even with singlelead AF ECGs, the proposed method estimated AAs accurately and calculated the atrial fibrillatory frequencies, the most valuable index in AF maintenance and therapy evaluation, with a remarkably low computational cost.
机译:本文提出了一种事件同步自适应滤波器(ESAF),用于从单导联AF ECG实时估计心房活动(AA)。拟议的ESAF是一种自适应滤波器,其设计是使脉冲序列馈入的参考信号与原始房颤(AF)ECG中的R峰值同步,并将及时延迟的AF ECG输入到主要输入中。为了评估性能,对于十个模拟AF心电图,计算了估计的AA与十个原始模拟AA之间的互相关系数(ρ)和归一化均方误差(NMSE),对于十个真实AF心电图,计算了心室残留(VR)在QRS间隔中计算相似度,在非QRS间隔中计算相似度(S)。结果,对于模拟的AF ECG,这四个参数显示为ρ= 0.938±0.016和NMSE = 0.243±0.051,对于真实的AF ECG,VR = 1.190±0.476和S = 0.967±0.041。发现这些结果优于以前被认为是实时自动自动估计AA的唯一方法的平均差减法(ABS)的结果。总之,即使使用单导联心电图心电图,所提出的方法也能准确估算AA并计算出房颤频率,这是房颤维持和治疗评估中最有价值的指标,并且计算成本非常低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号