首页> 外文期刊>Industrial Laboratory: Diagnostics of Materials >Atomic-absorption analysis used for determination of Ge and Bi in germanium and bismuth oxides and wastes in the production of bismuth ortho-germanate crystals
【24h】

Atomic-absorption analysis used for determination of Ge and Bi in germanium and bismuth oxides and wastes in the production of bismuth ortho-germanate crystals

机译:原子吸收分析法测定原锗酸铋晶体生产中锗和氧化铋及废物中的Ge和Bi

获取原文
获取原文并翻译 | 示例
           

摘要

Methods for atomic-absorption determination of germanium and bismuth in waste products of the production of bismuth ortho-germanate crystals are described. The relative standard deviations for bismuth analyzed in flame-atomized solutions (20-200 μg/ml) are 0.02 - 0.03 and for germanium (20 - 500 μg/ml)-0.015-0.04. Use of the electrothermal atomization techniques allows germanium and bismuth to be determined in the range 0.01 - 1.0 μg/m1.
机译:描述了原锗酸铋晶体生产中废品中锗和铋的原子吸收测定方法。在火焰雾化溶液中分析的铋(20-200μg/ ml)的相对标准偏差为0.02-0.03,对于锗(20-500μg/ ml)的相对标准偏差为-0.015-0.04。使用电热雾化技术可以将锗和铋的含量确定为0.01-1.0μg/ ml。