Brain biopsy performed with the RONNA G3 system: a case study on using a novel robotic navigation device for stereotactic neurosurgery

机译:用ronna g3系统进行的脑活动:使用新型机器人导航装置对立体定向神经外科的案检

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Background Robotic neuronavigation is becoming an important tool for neurosurgeons. We present a case study of a frameless stereotactic biopsy guided by the RONNA G3 robotic neuronavigation system. Methods A 45?year‐old patient with a history of vertigo, nausea and vomiting was diagnosed with multiple periventricular lesions. Neurological status was unremarkable. A frameless robotic biopsy of a brain lesion was performed. Results Three tissue samples were obtained. There were no intraoperative or postoperative complications. Histological analysis showed a B‐cell lymphoma. After merging the preoperative CT scan with the postoperative MRI and CT scans, the measured error between the planned and the postoperatively measured entry point was 2.24?mm and the measured error between the planned and postoperatively measured target point was 2.33?mm. Conclusions The RONNA G3 robotic system was used to navigate a Sedan brain biopsy needle to take tissue samples and could be a safe and precise tool for brain biopsy.
机译:抽象背景机器人神经元辐射成为神经外科医生的重要工具。我们展示了由ronna g3机器人神经元抗辐射系统引导的无框架立体定向活检的案例研究。方法45岁?历史患者患有眩晕,恶心和呕吐的历史,被诊断出患有多个脑盖状病变。神经系统状况不起眼。进行脑病变的无框架机器人活检。结果获得了三种组织样品。没有术中或术后并发症。组织学分析显示B细胞淋巴瘤。在利用术后MRI和CT扫描合并术前CT扫描之后,计划和垂直测量的入口点之间的测量误差为2.24Ωmm,计划和术后测量目标点之间的测量误差为2.33Ωmm。结论RONNA G3机器人系统用于导航轿车脑活检针进行组织样品,可以是脑活检的安全和精确的工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号