首页> 外文期刊>Topics in clinical nutrition >Comparison of Handheld to Metabolic Cart Indirect Calorimetry for Resting Energy Expenditure Assessment in Extremely Obese Women
【24h】

Comparison of Handheld to Metabolic Cart Indirect Calorimetry for Resting Energy Expenditure Assessment in Extremely Obese Women

机译:手持设备到代谢车间接量热法用于极端肥胖妇女的静息能量消耗评估的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Resting energy expenditure (REE) was measured by a Deltatrac (DT) metabolic cart and a MedGem (MG) handheld indirect calorimetry device in 14 extremely obese women. DT and MG REE measures were compared by Student's paired t test, correlation, and Bland-Altman plots. MG accuracy was also evaluated in comparison to DT by individual data inspection. DT and MG REE values were different on the basis of mean data. MG REE values were outside +-10% of DT REE in 6 of 14 subjects. It may be concluded from these data that the MG device may be unacceptable for clinical measurement of REE in this population.
机译:通过Deltatrac(DT)代谢推车和MedGem(MG)手持间接量热仪对14名极度肥胖的女性进行了静息能量消耗(REE)的测量。 DT和MG REE量度通过学生配对t检验,相关性和Bland-Altman图进行比较。与DT相比,MG准确性也通过单独的数据检查进行了评估。根据平均数据,DT和MG REE值不同。在14个受试者中的6个受试者中,MG REE值超出DT REE的+ -10%。从这些数据可以得出结论,MG设备对于该人群中REE的临床测量可能是不可接受的。