Contrast-enhanced MR imaging of the brain using T1-weighted FLAIR with BLADE compared with a conventional spin-echo sequence

机译:与传统的自旋回波序列相比,使用T1加权FLAIR和BLADE对比增强的大脑MR成像

获取原文
获取原文并翻译 | 示例

摘要

The BLADE and PROPELLER (periodically rotated overlapping parallel lines with enhanced reconstruction) techniques have been proposed to reduce the effect of head motion. Preliminary results have shown that BLADE also reduces pulsation artifacts from venous sinuses. The purpose of this study was to compare T1-weighted FLAIR acquired with BLADE (T1W-FLAIR BLADE) and T1-weighted spin-echo (T1W-SE) for the detection of contrast enhancement in a phantom and in patients with suspected brain lesions and to compare the degree of flow-related artifacts in the patients. A phantom filled with diluted Gd-DTPA was scanned in addition to 27 patients. In the phantom study, the peak contrast-to-noise ratio of T1W-FLAIR BLADE was larger than that of T1W-SE, and the position of the peak was shifted to a lower concentration. In patients, the degree of flow-related artifacts was significantly higher in T1W-SE. Among the 27 patients, 9 had metastatic tumor, and 18 did not. On a patient-by-patient basis, the sensitivity and specificity for the detection of metastatic lesions on axial T1W-SE were 100% and 55.6% respectively, while on axial T1W-FLAIR BLADE they were 100% and 100%. T1W-FLAIR BLADE seems to be capable of replacing T1W-SE, at least for axial post-contrast imaging to detect brain metastases.
机译:为了减少头部运动的影响,已经提出了BLADE和PROPELLER(周期性旋转的重叠平行线,具有增强的重构)技术。初步结果表明,BLADE还可以减少静脉窦产生的搏动伪影。这项研究的目的是比较用BLADE(T1W-FLAIR BLADE)和T1加权自旋回波(T1W-SE)获得的T1加权FLAIR,以检测幻影以及疑似脑损伤和比较患者中与流量相关的伪影的程度。除27例患者外,还扫描了装有稀释的Gd-DTPA的体模。在幻像研究中,T1W-FLAIR BLADE的峰值对比噪声比大于T1W-SE,并且峰值的位置移至较低的浓度。在患者中,T1W-SE中与血流有关的伪影程度明显更高。在27例患者中,有9例具有转移性肿瘤,而18例没有。在每个病人的基础上,在轴向T1W-SE上检测转移灶的敏感性和特异性分别为100%和55.6%,而在轴向T1W-FLAIR BLADE上检测转移灶的敏感性和特异性分别为100%和100%。 T1W-FLAIR BLADE似乎能够替代T1W-SE,至少可以用于轴向对比后成像以检测脑转移。

著录项

  • 来源
    《European Radiology》 |2008年第2期|337-342|共6页
  • 作者单位

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

    Department of Radiology Nagoya University Graduate School of Medicine 65 Tsurumai-cho Shouwa-ku Nagoya 466-8550 Japan;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    BLADE; Contrast; Flow; Metastasis; Magnetic resonance;

    机译:叶片;对比度;流量;转移;磁共振;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号