Performance of large-area avalanche photodiode for low-energy X-rays and y-rays scintillation detection

机译:大面积雪崩光电二极管在低能X射线和y射线闪烁检测中的性能

获取原文
获取原文并翻译 | 示例
           

摘要

We report on the performance of large-area (5 x 5 mm~2) avalanche photodiodes (APD) produced by Hamamatsu Photonics, as a low-energy X-rays and γ-rays scintillation detector. Hamamatsu APD has a reverse structure and works at relatively low bias voltage of 300-350 V. The leakage current is 1.2 nA at room temperature (25℃) and decreases to 10 pA at -20℃ for an avalanche gain of 50. The best FWHM energy resolutions of 9.4 +- 0.3% and 7.4 +- 0.3% were obtained for 59.5 keV y-rays from ~(241)Am and 122 keV γ-rays from ~(57)Co sources, respectively, as measured with a 5 x 5 x 5 mm~3 cubic CsI(Tl) crystal. We show that the minimum detectable energy for the scintillation light is remarkably low; 4.6 keV at room temperature (20℃) and 1.1 keV at -20℃. 5.9 keV X-rays from ~(55)Fe were clearly resolved at -20℃ with an FWHM resolution of 32.9 +- 0.3%. These results suggest that Hamamatsu APD can be a promising device for future applications in low-energy scintillation detection.
机译:我们报告了由Hamamatsu Photonics生产的大面积(5 x 5 mm〜2)雪崩光电二极管(APD)作为低能X射线和γ射线闪烁探测器的性能。滨松APD具有相反的结构,并在300-350 V的相对较低的偏置电压下工作。室温(25℃)下的泄漏电流为1.2 nA,在-20℃下的泄漏电流降至10 pA,雪崩增益为50。最佳用〜(5)测得的〜(241)Am的59.5 keV y射线和〜(57)Co的122 keVγ射线分别获得9.4±0.3%和7.4±0.3%的FWHM能量分辨率x 5 x 5 mm〜3立方CsI(Tl)晶体。我们表明,闪烁光的最小可检测能量非常低;在室温(20℃)下为4.6 keV,在-20℃下为1.1 keV。 〜(55)Fe产生的5.9 keV X射线在-20℃清晰分辨,FWHM分辨率为32.9±0.3%。这些结果表明,Hamamatsu APD可以作为低能耗闪烁检测未来应用的有前途的设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号