首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >A method for characterizing after-pulsing and dark noise of PMTs and SiPMs
【24h】

A method for characterizing after-pulsing and dark noise of PMTs and SiPMs

机译:表征PMT和SiPM的后脉冲和暗噪声的方法

获取原文
获取原文并翻译 | 示例

摘要

Photo-multiplier tubes (PMTs) and silicon photo-multipliers (SiPMs) are detectors sensitive to single photons that are widely used for the detection of scintillation and Cerenkov light in subatomic physics and medical imaging. This paper presents a method for characterizing two of the main noise sources that PMTs and SiPMs share: dark noise and correlated noise (after-pulsing). The proposed method allows for a model-independent measurement of the after-pulsing timing distribution and dark noise rate.
机译:光电倍增管(PMT)和硅光电倍增管(SiPM)是对单个光子敏感的检测器,广泛用于亚原子物理学和医学成像中闪烁和切伦科夫光的检测。本文提出了一种表征PMT和SiPM共享的两个主要噪声源的方法:暗噪声和相关噪声(后脉冲)。所提出的方法允许对后脉冲定时分布和暗噪声率进行模型无关的测量。