A method based on the coupling of a calixarene column used for the preconcentration of actinides to a quadrupole ICP-MS was developed for the separation of actinides and their on-line detection. First, a flow injection system was developed for aqueous U, Pu and Am on-line ICP-MS analysis. Then, chromatography parameters such as the column diameter and flow rates were optimised using the theoretical plate model. The detection limits achieved are 0.6 ng (7 μBq) for ~(238)U, 16 pg (37 mBq) for ~(239)Pu and 0.21 pg (27 mBq) for ~(241)Am. The signal improvement achieved by this coupling method was discussed and compared to "dilute and shoot" analysis. Finally, this method was validated on certified reference materials of ~(238)U, ~(239)Pu and ~(241)Am, at concentration levels close to the detection limits.
机译:开发了一种基于a化芳烃柱耦合的方法,用于cal系元素的预浓缩至四极杆ICP-MS,用于for系元素的分离和在线检测。首先,开发了一种流动注射系统,用于U,Pu和Am水溶液的在线ICP-MS分析。然后,使用理论塔板模型优化色谱参数,例如色谱柱直径和流速。达到的检测限为〜(238)U为0.6 ng(7μBq),〜(239)Pu为16 pg(37 mBq)和〜(241)Am为0.21 pg(27 mBq)。讨论了通过这种耦合方法实现的信号改善,并将其与“稀释并射击”分析进行了比较。最后,该方法在〜(238)U,〜(239)Pu和〜(241)Am的认证参考材料上进行了验证,浓度水平接近检测极限。

著录项

  • 来源
    《Journal of Analytical Atomic Spectrometry》 |2016年第7期|1406-1413|共8页
  • 作者单位

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France,PSL Research University, Chimie ParisTech - CNRS, Institut de Recherche de Chimie Paris, 11 rue Pierre et Marie Curie, 7S005 Paris, France;

    PSL Research University, Chimie ParisTech - CNRS, Institut de Recherche de Chimie Paris, 11 rue Pierre et Marie Curie, 7S005 Paris, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

    IRSN/Pole RadioProtection, Service de Dosimetrie Interne, LRC, 31, avenue de la division Leclerc, 92260 Fontenay-aux-Roses, France;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-18 03:04:01

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号