首页> 外文期刊>Pakistan journal of medical sciences. >Agreement between Keratometric readings by VERION image guided System, Galilei G4 and Pentacam
【24h】

Agreement between Keratometric readings by VERION image guided System, Galilei G4 and Pentacam

机译:VERION影像引导系统,Galilei G4和Pentacam的角膜测量读数之间的一致性

获取原文
           

摘要

Objective: To study Agreement between Keratometric readings by VERION image guided System, Galilei G4 and Pentacam. Methods: The quasi experimental study was conducted at Armed Forces Institute of Ophthalmology, Rawalpindi, Pakistan from August 2016 to December 2016. Twenty five patients fulfilling the inclusion criteria participated in the study. All Patients were subjected to Keratometric assessment using Galilei G4 Dual Scheimpflug analyzer (Ziemer, Switzerland), Wavelight Oculyzer II (Pentacam, Germany) and Verion image guided system (Alcon). Steep and flat meridian and diopter of astigmatism by three systems were recorded and endorsed. All readings were taken by the same observer. Statistical Program for Social Sciences (SPSS) version 22.0 was used for statistical analysis. Results analyzed for significance by t-tests and Interclass correlation analysis. In t tests, P values of 0.7 was considered acceptable. Results: Fifty eyes of twenty-five patients (22 male, 28 female) with mean age of 29.50 ± 3.46 years were studied. Flat K, steep K and dioptric power of astigmatism were measured with verion, pentacam and Galilei G4. Interclass correlation analysis showed agreement between individual variables measured by the three devices, while one sample t test showed no significant difference between dioptric power of astigmatism between Verion-Pentacam and Verion- Galilei group. (p 0.178 for former and 0.622 for later group). Conclusion: Verion image guided system is comparable to other instruments used currently for keratometry. Verion can be interchangeably used with Pentacam and Galilee G4 in assessing corneal astigmatism.
机译:目的:研究VERION图像引导系统,Galilei G4和Pentacam在角速度测量读数之间的一致性。方法:准实验研究于2016年8月至2016年12月在巴基斯坦拉瓦尔品第的武装部队眼科研究所进行。25名符合纳入标准的患者参加了该研究。使用Galilei G4 Dual Scheimpflug分析仪(瑞士Ziemer),Wavelight Oculyzer II(德国Pentacam)和Verion图像引导系统(Alcon)对所有患者进行角膜测量法评估。记录并认可了三个系统的陡峭平坦的子午线和屈光度。所有读数均由同一位观察员采集。使用社会科学统计程序(SPSS)22.0版进行统计分析。通过t检验和类间相关性分析对结果进行显着性分析。在t检验中,P值为0.7被认为是可以接受的。结果:研究了25例平均年龄为29.50±3.46岁的患者(22例男性,28例女性)的眼睛。使用verion,pentacam和Galilei G4测量了平坦K,陡峭K和散光度数。类间相关性分析表明,使用这三种装置测量的各个变量之间具有一致性,而一个样本t检验显示,Verion-Pentacam和Verion-Galilei组的散光屈光力之间无显着差异。 (前者p为0.178,后者为0.622)。结论:Verion图像引导系统可与当前用于角膜曲率测定的其他仪器相媲美。 Verion可与Pentacam和Galilee G4互换使用,以评估角膜散光。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号