首页> 外文期刊>International ophthalmology >Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer
【24h】

Comparison of keratometric measurements obtained by the Verion Image Guided System with optical biometry and auto-keratorefractometer

机译:使用光学生物学和自动角划晕器的VERION图像引导系统获得的静脉图像导向系统比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The aim of this study was to compare the keratometric measurements of Verion Image Guided System with an optical biometer (Zeiss IOLMaster 500, Carl Zeiss Meditec, Jena, Germany) and an automated keratorefractometer (AKR) (Topcon KR-8900, Topcon, Japan). In this prospective clinical trial, the right eyes of 52 patients with cataract were examined (mean age 62.25 ± 12.16 years). The measurements were taken by the three systems in a random order. Keratometric data, magnitude of astigmatism, and astigmatic axis measurements from all three instruments were compared. The results were evaluated using, intraclass correlation coefficients (ICC), Bland-Altman plots, and paired samples t tests. The mean flat/steep K of Verion, IOLMaster, and AKR were 43.22 ± 1.38D/44.23 ± 1.46D, 43.07 ± 1.26D/44.05 ± 1.34D, and 43.07 ± 1.31D/43.89 ± 1.42D, respectively. Flat K readings of Verion were higher than IOLMaster and AKR (p 0.05 for both). Steep K readings were different for all three (p 0.05). The magnitude of astigmatism by Verion and IOLMaster were 0.98 ± 0.65D and 0.98 ± 0.59D (p = 0.88). The mean astigmatism measured by the AKR was 0.82 ± 0.62D, less than the other two instruments (p 0.001). Astigmatic axis measurements of Verion and AKR differed <10° in 38, between 10° and 20° in 5, and >20° in 9 eyes; the same difference was 30, 11, and 11 eyes, respectively, between Verion and IOLMaster. Although, keratometric and astigmatic results obtained from Verion were not completely interchangeable with IOLMaster and AKR, especially the agreement between Verion and IOLMaster was excellent with ICCs close to one. However, there were pronounced astigmatic axis measurement differences between three instruments.
机译:本研究的目的是将Verion图像引导系统的静脉测量测量与光学生物计(Zeiss Iolmaster 500,Carl Zeiss Meditec,Jena,德国)和一辆自动的角鲨(AKR)(Topcon Kr-8900,Topcon,Japan)进行了比较。在这项前瞻性临床试验中,检查了52例白内障患者的右眼(平均年龄62.25±12.16岁)。测量由三个系统以随机顺序拍摄。比较了所有三种仪器的静脉内数据,散光幅度和散光轴测量。使用脑内相关系数(ICC),Bland-Altman图和配对样本T测试评估结果。 VERION,IOLMASTER和AKR的平均扁平/陡峭K分别为43.22±1.38d / 44.23±1.46d,43.07±1.26d / 44.05±1.34d和43.07±1.31d / 43.89±1.42d。普通的k读数读数高于iolmaster和akr(两者的p 0.05)。所有三种(P 0.05)的陡峭K读数不同Verion和Iolmaster的散光幅度为0.98±0.65d和0.98±0.59d(p = 0.88)。通过AKR测量的平均散光度为0.82±0.62D,小于其他两种仪器(P 0.001)。 Arigmatic轴测量的验尸和Akr在38中的58分,在5°和20°之间,在9只眼中为20°;在Verion和Iolmaster之间,相同的差异分别为30,11和11只眼睛。尽管从妊娠中获得的静脉测定和散光结果并不完全可与IOLMASTER和AKR互换,但特别是验证和IOLMASTER之间的协议与靠近一个的ICCS非常好。但是,在三种仪器之间存在明显的散光轴测量差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号