首页> 外文期刊>IEEE Transactions on Biomedical Engineering >Design and initial evaluation of an implantable sonomicrometer and CW Doppler flowmeter for simultaneous recordings with a multichannel telemetry system
【24h】

Design and initial evaluation of an implantable sonomicrometer and CW Doppler flowmeter for simultaneous recordings with a multichannel telemetry system

机译:用于多通道遥测系统同时记录的植入式体温计和CW多普勒流量计的设计和初步评估

获取原文
获取原文并翻译 | 示例

摘要

We have developed a sonomicrometer and continuous wavelength (CW) Doppler flowmeter for a multichannel telemetry system. These developments will enable us to measure ventricular dimension and coronary artery blood velocity, which are valuable parameters to characterize sudden cardiac death (SCD) in ambulatory animal models of ventricular arrhythmias. The design goals for the sensors were accuracy, low power consumption, small size and compatibility with each other. The circuits were designed successfully and tested simultaneously in vivo. The CW Doppler flowmeter draws 9 mA and the sonomicrometer draws 28 mA on a 5-V supply. The ability to measure heart dimension and blood velocity will add significantly to our understanding of the sequence of events leading up to spontaneous sudden cardiac death.
机译:我们已经开发了用于多通道遥测系统的体视计和连续波长(CW)多普勒流量计。这些发展将使我们能够测量心室尺寸和冠状动脉血流速度,这是表征室性心律不齐非卧床动物模型中突发性心脏死亡(SCD)的有价值的参数。传感器的设计目标是精度,低功耗,小尺寸和彼此兼容。电路设计成功,并在体内同时进行了测试。在5 V电源上,CW多普勒流量计消耗9 mA电流,而电量计消耗28 mA。测量心脏尺寸和血流速度的能力将大大增加我们对导致自发性心源性猝死的事件顺序的理解。