首页> 外文期刊>IEEE Transactions on Applied Superconductivity >Experimental apparatus for critical current measurement above 5 Kusing Bi-based oxide current leads
【24h】

Experimental apparatus for critical current measurement above 5 Kusing Bi-based oxide current leads

机译:用于测量超过5个使用Bi基氧化物电流引线的临界电流的实验设备

获取原文
获取原文并翻译 | 示例

摘要

Experimental apparatus for the measurement of temperature dependence of critical current within a strong magnetic field has been developed. Samples, which are tested by the four probe method are cooled by a Gifford-McMahon (GM) type cryocooler to the lowest temperature of 5 K. The sample holder has a diameter of 70 mm and a height of 70 mm. Transport currents, up to 150 A, are supplied to two samples through Bi-based oxide superconducting current leads. An external magnetic field of up to 15 T, which is applied to the samples perpendicular to the sample axis, is generated by a water cooled single Bitter magnet with a room temperature bore of 82 mm. As a performance test, the critical current properties of a silver clad Bi-based oxide tape have been measured at 5.5 K, 30 K and 50 K within a magnetic field up to 15 T. This experiment has demonstrated the validity of this experimental apparatus
机译:已经开发出用于测量强磁场内的临界电流的温度依赖性的实验装置。通过四探针法测试的样品通过Gifford-McMahon(GM)型低温冷却器冷却至最低温度5K。样品架的直径为70 mm,高度为70 mm。通过Bi基氧化物超导电流引线将高达150 A的传输电流提供给两个样品。垂直于样品轴施加到样品的高达15 T的外部磁场是由具有82 mm室温孔的水冷单个Bitter磁体产生的。作为一项性能测试,已在高达15 T的磁场中在5.5 K,30 K和50 K的条件下测量了包覆银的Bi基氧化带的临界电流性能。该实验证明了该实验装置的有效性