Magnetic localization and control of helical robots for clearing superficial blood clots

机译:螺旋机器人的磁性定位和控制用于清除浅层血块

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This work presents an approach for the localization and control of helical robots during removal of superficial blood clots inside in vitro and ex vivo models. The position of the helical robot is estimated using an array of Hall-effect sensors and precalculated magnetic field map of two synchronized rotating dipole fields. The estimated position is used to implement closed-loop motion control of the helical robot using the rotating dipole fields. We validate the localization accuracy by visual feedback and feature tracking inside the in vitro model. The experimental results show that the magnetic localization of a helical robot with diameter of 1 mm can achieve a mean absolute position error of 2.35 ± 0.4 mm (n = 20). The simultaneous localization and motion control of the helical robot enables propulsion toward a blood clot and clearing at an average removal rate of 0.67 ± 0.47 mm3/min. This method is used to localize the helical robot inside a rabbit aorta (ex vivo model), and the localization accuracy is validated using ultrasound feedback with a mean absolute position error of 2.6 mm.
机译:这项工作提出了一种在体外和离体模型内去除浅层血块期间对螺旋机器人进行定位和控制的方法。使用一系列霍尔效应传感器和两个同步旋转偶极子场的预先计算的磁场图来估计螺旋机器人的位置。估计的位置用于使用旋转偶极子场对螺旋机器人进行闭环运动控制。我们通过视觉反馈和体外模型内部的特征跟踪来验证定位精度。实验结果表明,直径为1 mm的螺旋机器人的磁定位可以实现2.35 absolute±error0.4 mm的平均绝对位置误差(n = 20)。螺旋机器人的同时定位和运动控制,可将血栓推进并清除,平均清除速度为0.67 average±0.47 mm 3 / min。该方法用于在兔主动脉(离体模型)中定位螺旋机器人,并且使用超声反馈(平均绝对位置误差为2.6mm)验证了定位精度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号