The design and construction of an oxygen micro-electrode for monitoring 20-100 microliter volumes.

机译:用于监测20-100微升体积的氧气微电极的设计和构造。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Full instructions are given for the construction of a micro-respirometer. It is based on a membrane-covered platinum electrode of 10 micron diameter that consumes little oxygen. The reaction chamber, made entirely of glass, can be varied in volume between 20 and 100 microliter by adjusting the sealer assembly. Although specifically designed to study the properties of small samples of mitochondria from tiny organisms, it can be used equally well either for oxygen dissolved in aqueous solutions or dry gas mixtures containing oxygen. Measurements can be made with about one-hundredth the amount of material required for commercially available macro-respirometers, and when connected via an impedance-matching device to a pen recorder it gives a stable output of 100-200 mV in air.
机译:给出了有关微呼吸计的构造的完整说明。它基于10微米直径的膜覆盖铂电极,消耗很少的氧气。完全由玻璃制成的反应腔室,可以通过调节密封组件来在20到100微升之间变化。尽管专门设计用于研究来自微小生物体的线粒体小样本的特性,但它对于溶解在水溶液中的氧气或含有氧气的干燥气体混合物同样适用。可以用市售的大型呼吸计所需材料的大约一百倍进行测量,当通过阻抗匹配设备连接到笔式记录仪时,它在空气中的输出稳定为100-200 mV。