Reproducibility and accuracy of measurements with a hand held autorefractor in children

机译:儿童手持自动验光仪的测量重现性和准确性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

AIM—To determine reproducibility and accuracy of the Nikon Retinomax autorefractor when used with children who were made cycloplegic.
METHODS—Autorefraction and retinoscopy or subjectively refined retinoscopy (where, under the patient's direction, the refraction was varied until the best visual acuity was achieved) were performed on the right eye of 47 children, age 11-93 months. Autorefraction was performed using the Nikon Retinomax, which provides up to eight measured values of refractive error and one representative measurement of refractive error.
RESULTS—Autorefractor measurements were successfully obtained from 7/9 children age 3 years or younger, and from all older children. Vector methods were used to calculate differences. Retinomax reproducibility averaged 0.43 D. Unbiased Retinomax and retinoscopy measurements differed by an average of 0.82 D. Unbiased Retinomax and subjectively refined retinoscopy differed by an average of 1.03 D.
CONCLUSIONS—Reproducibility of Retinomax measured values in children is comparable with reproducibility of retino scopy, subjective refraction, and autorefraction measurements in adults. Agreement between Retinomax and retinoscopy and agreement between Retinomax and subjective refinement in children is comparable with agreement between autorefraction and subjective refraction in adults. The study indicates that the Retinomax is a useful instrument for measuring refractive errors in young children.

机译:目的:确定尼康Retinomax自动验光仪与患有睫状肌麻痹的儿童一起使用时的可重复性和准确性。已对47位年龄在11-93个月的儿童的右眼进行了矫正。使用尼康Retinomax进行自动验光,可提供多达8个屈光不正的测量值和一个代表性的屈光不正的测量结果。
结果-从3/7岁或以下的7/9儿童中成功获得了自动验光仪的测量结果所有大孩子。使用向量法计算差异。 Retinomax的再现性平均为0.43 D.无偏的Retinomax和检影法的平均差异为0.82 D.无偏的Retinomax和主观精巧的检影法的平均差异为1.03 D。在儿童中,该值可与成年人的视网膜镜检,主观验光和自动验光测量的再现性相媲美。 Retinomax与检影镜之间的一致性以及Retinomax与儿童主观精细化之间的一致性可与成人的自动验光与主观屈光之间的一致性相提并论。研究表明,Retinomax是一种用于测量幼儿屈光不正的有用仪器。