Gas Flowrate Calibration Using a Piston Prover and A Laser Doppler Displacement Meter

机译:使用活塞证报和激光多普勒位移计的气体流量校准

获取原文

摘要

A Laser Doppler Displacement Meter (LDDM) is used to improve the performance of a mercury sealed piston device for the calibration of gas flow at low flow rates. With a 0.5 inch retroreflector at the center of the piston and a LDDM laser head at the top of the tube, the piston displacement and its corresponding velocity can be measured with a resolution down to 10 nm and a data rate of up to 800,000 data/sec. The advantage of using the LDDM is fast and accurate readings not affected by piston rocking or sticking. Preliminary test data have shown an accuracy of 0.2% of reading. Based on an error budget calculation, an accuracy of 0.1% of reading can be readily achieved with better pressure and temperature sensors.
机译:激光多普勒位移计(LDDM)用于改善汞密封活塞装置的性能,以便以低流速校准气体流量。在活塞的中心处的0.5英寸的后向图和管顶部的LDDM激光头处,可以通过降至10nm的分辨率测量活塞位移及其相应的速度,数据速率高达800,000个数据/秒。使用LDDM的优点是不受活塞摇摆或粘附影响的快速准确的读数。初步测试数据显示了读数的0.2%的精度。基于错误预算计算,可以随着更好的压力和温度传感器容易地实现0.1%的准确度。