首页> 外文期刊>Measurement Science & Technology >The calibration of a turbine meter as a secondary flow-rate standard by a one-point relative method using laser Doppler velocimetry
【24h】

The calibration of a turbine meter as a secondary flow-rate standard by a one-point relative method using laser Doppler velocimetry

机译:使用激光多普勒测速仪通过单点相对方法将涡轮流量计作为二级流量标准进行校准

获取原文
获取原文并翻译 | 示例
           

摘要

Heat district suppliers demand carefully constructed and calibrated flow sensors of heat meters at temperatures above 100 ℃ for long-term use. The calibration at these temperatures cannot be performed by the common gravimetric standard method due to technical difficulties. A novel one-point relative method allows a laser Doppler velocimeter (LDV) was therefore developed which is based on the gravimetric method below 100℃ and on the Reynolds law. This method allows a turbine meter to be calibrated as a secondary flow-rate standard with a relative uncertainty #0.2% for temperatures up to 180 ℃ with regard to the flow rate.
机译:供热区的供应商要求精心设计和校准的热表流量传感器在100℃以上的温度下才能长期使用。由于技术上的困难,无法通过常规的重量标准方法在这些温度下进行校准。因此,基于低于100℃的重量法和雷诺定律,开发了一种新颖的单点相对方法允许激光多普勒测速仪(LDV)。这种方法允许将涡轮流量计校准为次级流量标准,对于高达180℃的温度,其相对不确定度为#0.2%。