首页> 外文会议>International Conference on Infrared, Millimeter, and Terahertz Waves >Development of Millimeter-Wave Fabry-Perot Resonator for Simultaneous Electron-Spin and Nuclear-Magnetic Resonance Measurement at Low Temperatures
【24h】

Development of Millimeter-Wave Fabry-Perot Resonator for Simultaneous Electron-Spin and Nuclear-Magnetic Resonance Measurement at Low Temperatures

机译:毫米波法布里-珀罗共振器在低温下同时进行电子自旋和核磁共振测量的开发

获取原文

摘要

We have developed millimeter-wave Fabry-Perot type resonator (FPR) aiming for simultaneous electron-spin and nuclear-spin resonance (ESR/NMR) measurements at very low temperatures below 1 K. The flat mirror of FPR is fabricated using a gold thin layer in order to minimize disturbance to radio frequency wave for NMR. An optimum thickness of the gold layer by measuring Q value is found. We have actually succeeded to detect both ESR and NMR at very low temperatures.
机译:我们开发了毫米波Fabry-Perot型谐振器(FPR),旨在在低于1 K的极低温度下同时进行电子自旋和核自旋共振(ESR / NMR)测量。为了最大程度地减小对NMR的射频波的干扰。通过测量Q值找到金层的最佳厚度。实际上,我们已经成功地在非常低的温度下检测了ESR和NMR。