首页> 外文会议>International Measurement Federation world congress >DEVELOPMENT OF METHOD FOR DETERMINATION OF PRESSURE BALANCE PISTON FALL RATE
【24h】

DEVELOPMENT OF METHOD FOR DETERMINATION OF PRESSURE BALANCE PISTON FALL RATE

机译:测定压力平衡活塞降速方法的开发

获取原文

摘要

This paper describes internal method for determination of pressure balance piston fall rate using simple camera-based optical system with laboratory developed software. Measurements were carried out on three standard piston/cylinder units in Croatian National Pressure Laboratory (LPM) using gas and oil as transmitting medium. Measurement equipment, procedure and fall rate results for three set of measurements are given as well as evaluation of measurement uncertainty. Results were compared with other relevant measurements.
机译:本文介绍了使用简单的基于摄像头的光学系统以及实验室开发的软件来确定压力平衡活塞下降率的内部方法。在克罗地亚国家压力实验室(LPM)中使用气体和石油作为传输介质,在三个标准活塞/气缸单元上进行了测量。给出了三组测量的测量设备,过程和下降率结果,以及对测量不确定度的评估。将结果与其他相关测量结果进行比较。