This paper describes internal method for determination of pressure balance piston fall rate using simple camera-based optical system with laboratory developed software. Measurements were carried out on three standard piston/cylinder units in Croatian National Pressure Laboratory (LPM) using gas and oil as transmitting medium. Measurement equipment, procedure and fall rate results for three set of measurements are given as well as evaluation of measurement uncertainty. Results were compared with other relevant measurements.
机译:本文介绍了使用实验室开发软件的简单相机的光学系统确定压力平衡活塞率的内部方法。使用气体和油作为传输介质,在克罗地亚国家压力实验室(LPM)的三个标准活塞/气缸装置上进行测量。给出了三组测量的测量设备,程序和跌倒结果以及测量不确定性的评估。结果与其他相关测量进行了比较。