首页> 外文会议>19th ITG/GMA-Symposium Sensors and Measuring Systems >Multicomponent force/torque sensor with integrated calibration system
【24h】

Multicomponent force/torque sensor with integrated calibration system

机译:带有集成校准系统的多分量力/扭矩传感器

获取原文
获取原文并翻译 | 示例

摘要

A multicomponent force and torque sensor with integrated calibration capabilities is shown. The system uses the Kibblebalance principle to provide traceable force and torque measurements. This is possible by integrating standards for position, angle, voltage, resistance and time measurement. In this way, calibration operations based on deadweights and the use of local gravity acceleration are avoided. A prototype of the system for measuring forces and torques in a range of +-2 N and +-0.1 N m respectively was implemented. A maximal relative expanded measurement uncertainty (k=2) of 1.10(exp -4) is expected for the force and torque measurements. Experimental results used to calibrate the integrated position/ angle and voltage measuring systems are shown.
机译:显示了具有集成校准功能的多分量力和扭矩传感器。该系统使用Kibblebalance原理提供可追溯的力和扭矩测量。通过集成用于位置,角度,电压,电阻和时间测量的标准,这是可能的。以这种方式,避免了基于自重的校准操作和局部重力加速度的使用。实现了用于分别在+ -2 N和+ -0.1 N m范围内测量力和扭矩的系统原型。力和扭矩测量的最大相对扩展测量不确定度(k = 2)为1.10(exp -4)。显示了用于校准集成位置/角度和电压测量系统的实验结果。