首页> 外文OA文献 >Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates
【2h】

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

机译:手术扭矩限制器的有效性:比较钻头和手动螺钉插入锁定板的模型

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background: The objective of this study is to analyse the effectiveness of the surgical torque limiter during operative use. The study also investigates the potential differences in torque between hand and drill-based screw insertion into locking plates using a standardised torque limiter. Methods: Torque for both hand and power screw insertion was measured through a load cell, registering 6.66 points per second. This was performed in a controlled environment using synthetic bone, a locking plate and locking screws to simulate plate fixation. Screws were inserted by hand and by drill with torque values measured. Results: The surgical torque limiter (1.5 Nm) was effective as the highest recorded reading in the study was 1.409 Nm. Comparatively, there is a statistically significant difference between screw insertion methods. Torque produced for manually driven screw insertion into locking plates was 1.289 Nm (95 % CI 1.269–1.308) with drill-powered screw insertion at 0.740 Nm (95 % CI 0.723–0.757). Conclusions: The surgical torque limiter proved to be effective as per product specifications. Screws inserted under power produce significantly less torque when compared to manual insertion by hand. This is likely related to the mechanism of the torque limiter when being used at higher speeds for which it was designed. We conclude that screws may be inserted using power to the plate with the addition of a torque limiter. It is recommended that all screws inserted by drill be hand tightened to achieve adequate torque values.
机译:背景:本研究的目的是分析手术期间使用扭矩限制器的有效性。这项研究还研究了使用标准扭矩限制器将手和基于钻的螺钉插入锁定板之间在扭矩方面的潜在差异。方法:通过称重传感器测量手动和动力螺钉插入的扭矩,每秒记录6.66点。这是在受控环境中使用合成骨,锁定板和锁定螺钉来模拟板固定的操作。用手和钻头插入螺钉,并测量扭矩值。结果:手术扭矩限制器(1.5 Nm)有效,因为该研究中记录的最高读数为1.409 Nm。相比之下,螺钉插入方法之间在统计学上有显着差异。手动将螺钉插入锁紧板时产生的扭矩为1.289 Nm(95%CI 1.269–1.308),而钻头动力螺钉插入时的扭矩为0.740 Nm(95%CI 0.723–0.757)。结论:手术扭矩限制器被证明是有效的,根据产品规格。与手动插入相比,通电插入的螺钉产生的扭矩要小得多。这可能与扭矩限制器的设计有关,该限制器在以较高速度设计时使用。我们得出的结论是,可以在增加扭矩限制器的情况下使用动力将螺钉插入板上。建议手动拧紧钻头插入的所有螺钉,以获得足够的扭矩值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号