首页> 外文OA文献 >Fontaine atomique double de césium et de rubidium avec une exactitude de quelques 10^-16 et applications
【2h】

Fontaine atomique double de césium et de rubidium avec une exactitude de quelques 10^-16 et applications

机译:铯和rub双原子喷泉(精度约为10 ^ -16)及其应用

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Atomic fountains are the most accomplished development of the atomic clocks based on the cesium atom whose hyperfine resonance defines the SI second since 1967. Today these systems are among those which realize the second with the best accuracy. We present the last developments of the cold cesium and rubidium atom dual fountain experiment at LNE-SYRTE. This unique dual setup would allow to obtain an outstanding resolution in fundamental physics tests based on atomic transition frequency comparisons. In order to enable operation with both atomic species simultaneously, we designed, tested and implemented on the fountain new collimators which combine the laser lights corresponding to each atom. By comparing our rubidium fountain to another cesium fountain over a decade, we performed a test of the stability of the fine structure constant at the level of 5E-16 per year. We carried on the work on the clock accuracy and we focused on the phase gradients effects in the interrogation cavity and on the microwave leakage. The fountain accuracy has been evaluated to 4E-16 for the cesium clock and to 5E-16 for the refurbished rubidium clock. As a powerful instrument of metrology, our fountain was implicated in many clock comparisons and contributed many times to calibrate the International Atomic Time. Furthermore, we used the fountain to perform a new test of Lorentz local invariance.
机译:原子喷泉是最先进的基于铯原子的原子钟的开发,铯原子的超精细共振自1967年以来就定义了SI秒。如今,这些系统属于那些以最高精度实现秒的系统。我们介绍了LNE-SYRTE上冷铯和rub原子双喷头实验的最新进展。这种独特的双重设置可以在基于原子跃迁频率比较的基础物理测试中获得出色的分辨率。为了使两种原子同时运行,我们在喷泉上设计,测试和实现了新的准直器,这些准直器将与每个原子相对应的激光组合在一起。通过将我们的id喷泉与另一个铯喷泉进行比较,我们对精细结构常数(每年5E-16的水平)的稳定性进行了测试。我们对时钟精度进行了研究,重点研究了讯问腔中的相位梯度效应和微波泄漏。铯钟的喷泉精度已评估为4E-16,翻新rub钟的喷泉精度已评估为5E-16。作为一种功能强大的计量仪器,我们的喷泉涉及许多时钟比较,并且为校准国际原子钟做出了许多贡献。此外,我们使用了源泉对洛伦兹局部不变性进行了新的检验。

著录项

  • 作者

    Chapelet Frédéric;

  • 作者单位
  • 年度 2008
  • 总页数
  • 原文格式 PDF
  • 正文语种 fr
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号