首页> 美国政府科技报告 >High-Frequency Percussive Ventilation: Pneumotachograph Validation and Tidal Volume Analysis
【24h】

High-Frequency Percussive Ventilation: Pneumotachograph Validation and Tidal Volume Analysis

机译:高频冲击通气:肺气管造影检查和潮气量分析

获取原文

摘要

INTRODUCTION: High-frequency percussive ventilation (HFPV) is an increasingly used mode of mechanical ventilation, for which there is no proven real-time means of measuring delivered tidal volume (VT). OBJECTIVE: To validate a pneumotachograph for HFPV and then exploit flowsensor data to describe the behavior of both low-frequency and high-frequency breaths. METHODS: Sensor performance was gauged during changes in high-frequency and low- frequency rate and ratio, mean airway pressure, oxygen concentration, heated or heated-humidified gas flow, and endotracheal tube diameter. Glass bottle (adiabatic VT) and test lung (adiabatically derived low-frequency VT) based adiabatic conditions provided both an initial source for analog signal calibration and an accepted standard comparator to flow-sensor measurement of high-frequency and low-frequency (flow-sensor-derived) VT, respectively. RESULTS: Pneumotachography proved accurate and precise over an array of tested settings and conditions when analyzing both high-frequency and low-frequency breaths.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号