首页> 外文期刊>Neurourology and urodynamics. >Evaluating the use of different waveforms for intravesical electrical stimulation: a study in the rat.
【24h】

Evaluating the use of different waveforms for intravesical electrical stimulation: a study in the rat.

机译:评价不同波形对膀胱内电刺激的使用:一项在大鼠中的研究。

获取原文
获取原文并翻译 | 示例

摘要

AIM: For decades, monophasic square wave pulses have been used for intravesical electrical stimulation (IVES) in diagnosis and treatment of lower urinary tract dysfunction. The use of other waveforms, more preferred by patients or more effective for neural stimulation, has hardly been studied before. This study compares the effect of several waveforms and derived parameters in IVES. METHODS: IVES, performed in eight female Sprague-Dawley rats, was given for 300 sec with constant frequency (10 Hz) and current amplitude (6 mA) with unipolar square, biphasic square, asymmetric biphasic square, double square, unipolar exponential rise, biphasic exponential rise, and double-exponential rise wavepulses with different pulse durations (2 x 5 msec, 20 msec, or 2 x 20 msec). Bladder pressure was recorded. Average current and average power were calculated. RESULTS: Stimulations with the different waveforms resulted in the same maximal pressure rise (all P>0.1). The average charge was zero for biphasic waveforms. For exponential rise pulse waveforms the total injected charge was lower than that for the monophasic square pulse waveform. The average power was the highest for biphasic waveforms and the lowest for monophasic double-exponential rise pulses waveform. CONCLUSIONS: Charge-balanced waveforms, which are more preferred by patients, are equally effective in evoking IVES-induced detrusor contraction as monophasic square pulse waveforms. Thus, it can be beneficial in clinical practice to replace the monophasic square pulse waveform for IVES by charge-balanced waveforms. Furthermore, the stimulation power and thus the requested energy can be reduced by changing the waveform. This can be important for electrical stimulation using implanted batteries.
机译:目的:几十年来,单相方波脉冲已用于膀胱内电刺激(IVES)以诊断和治疗下尿路功能障碍。以前几乎没有研究过使用其他更受患者青睐或对神经刺激更有效的波形。这项研究比较了IVES中几个波形和派生参数的影响。方法:对八只雌性Sprague-Dawley大鼠进行IVES,持续300秒,以恒定频率(10 Hz)和电流幅度(6 mA)进行,单极方,双相方,不对称双相方,双方,单极指数上升,双相指数上升和具有不同脉冲持续时间(2 x 5毫秒,20毫秒或2 x 20毫秒)的双指数上升波脉冲。记录膀胱压力。计算了平均电流和平均功率。结果:不同波形的刺激导致相同的最大压力上升(所有P> 0.1)。双相波形的平均电荷为零。对于指数上升脉冲波形,总注入电荷低于单相方波波形。对于双相波形,平均功率最高,而对于单相双指数上升脉冲波形,平均功率最低。结论:电荷平衡波形,更受患者青睐,在诱发IVES诱导的逼尿肌收缩方面与单相方波波形同样有效。因此,在临床实践中用电荷平衡波形代替IVES的单相方波波形可能是有益的。此外,可以通过改变波形来减小刺激功率并因此减小所需能量。这对于使用植入式电池进行电刺激可能很重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号