An amperometric glucose biosensor based on a screen-printed electrode and Os-complex mediator for flow injection analysis

机译:基于丝网印刷电极和Os复杂介体的电流型葡萄糖生物传感器用于流动注射分析

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

An amperometric Os-complex mediated glucose biosensor based on a three electrodes planar configuration was constructed using screen printing technique for flow injection analysis. The electrochemical characteristic of the amperometric biosensor was evaluated by the cyclic voltammetry in the absence and presence of Os-complex mediator. The Os-complex mediated biosensor was found to have a C-V curve close to an ideal. The applied potential of the biosensor had been explored and -0.1 V vs. Ag|AgCl reference electrode was selected for avoiding excessive interference from other electroactive species. The biosensor also indicated good stability of more than 90percent of its original activity for about 14 months stored at 4 deg C. The biosensor on the detection of glucose showed a linear response range of 0.5-30 mM and the correlation coefficient up to 0.99 together with a FIA system. The coefficient of variation was 3.3percent for injections with a single biosensor (n velence 12). The results show that the biosensor has fast response, good sensitivity and reproducibility for flow injection analysis.
机译:使用丝网印刷技术构建了基于三电极平面构型的安培型Os络合物介导的葡萄糖生物传感器,用于流动注射分析。在不存在Os-复合物介体的情况下,通过循环伏安法评估电流型生物传感器的电化学特性。发现Os复合物介导的生物传感器的C-V曲线接近理想值。已经探索了生物传感器的施加电势,并选择了-0.1 V vs. Ag | AgCl参比电极,以避免来自其他电活性物质的过度干扰。该生物传感器还显示出在4℃下储存约14个月时其原始活性的90%以上具有良好的稳定性。该生物传感器在检测葡萄糖时显示出0.5-30 mM的线性响应范围,相关系数高达0.99,并且国际汽联系统。使用单个生物传感器进样的变异系数为3.3%(nvelence 12)。结果表明,该生物传感器具有快速响应,良好的灵敏度和重现性,可用于流动注射分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号