Gamma knife output factor measurements using VIP polymer gel dosimetry.

机译:使用VIP聚合物凝胶剂量法测量伽马刀输出因子。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Water equivalent polymer gel dosimeters and magnetic resonance imaging were employed to measure the output factors of the two smallest treatment fields available in a Gamma Knife model C radiosurgery unit, those formed employing the 4 and 8 mm final collimator helmets. METHODS: Three samples of the VIP normoxic gel formulation were prepared and irradiated so that a single shot of the field whose output factor is to be measured and a single shot of the reference 18 mm field were delivered in each one. Emphasis is given to the development and benchmarking of a refined data processing methodology of reduced uncertainty that fully exploits the 3D dose distributions registered in the dosimeters. RESULTS: Polymer gel results for the output factor of the 8 mm collimator helmet are found to be in close agreement with the corresponding value recommended by the vendor (0.955 +/- 0.007 versus 0.956, respectively). For the 4 mm collimator helmet, however, polymer gel results suggest an output factor 3% lower than the value recommended by the vendor (0.841 +/- 0.009 versus 0.870, respectively). CONCLUSIONS: A comparison with corresponding measurements published in the literature indicates that output factor results of this work are in agreement with those obtained using dosimetric systems which, besides fine spatial resolution and lack of angular and dose rate dependence of the dosimeter's response, share with polymer gels the favorable characteristic of minimal radiation field perturbation.
机译:目的:使用等效水凝胶凝胶剂量计和磁共振成像技术测量Gamma Knife C型放射外科设备中可用的两个最小治疗区域的输出因子,这些区域使用4毫米和8毫米最终准直仪头盔形成。方法:准备并辐照了VIP常氧凝胶制剂的三个样品,以使每一个都可以测量其输出因子的场的单次发射和参考18 mm场的单次发射。重点在于减少不确定性的完善数据处理方法的开发和基准测试,该方法可充分利用剂量计中记录的3D剂量分布。结果:发现8毫米准直仪头盔的输出因子的聚合物凝胶结果与供应商推荐的相应值非常一致(分别为0.955 +/- 0.007和0.956)。但是,对于4 mm的准直仪头盔,聚合物凝胶的结果表明输出因子比供应商建议的值低3%(分别为0.841 +/- 0.009和0.870)。结论:与文献中发表的相应测量结果进行比较表明,这项工作的输出因子结果与使用剂量学系统获得的结果一致,该剂量学系统除了具有良好的空间分辨率以及剂量计响应的角度和剂量率依赖性外,还与聚合物共享凝胶化了最小的辐射场扰动的有利特性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号