Gamma knife output factor measurements using VIP polymer gel dosimetry.

机译:使用VIP聚合物凝胶剂量测定法测量伽马刀输出因子测量。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Water equivalent polymer gel dosimeters and magnetic resonance imaging were employed to measure the output factors of the two smallest treatment fields available in a Gamma Knife model C radiosurgery unit, those formed employing the 4 and 8 mm final collimator helmets. METHODS: Three samples of the VIP normoxic gel formulation were prepared and irradiated so that a single shot of the field whose output factor is to be measured and a single shot of the reference 18 mm field were delivered in each one. Emphasis is given to the development and benchmarking of a refined data processing methodology of reduced uncertainty that fully exploits the 3D dose distributions registered in the dosimeters. RESULTS: Polymer gel results for the output factor of the 8 mm collimator helmet are found to be in close agreement with the corresponding value recommended by the vendor (0.955 +/- 0.007 versus 0.956, respectively). For the 4 mm collimator helmet, however, polymer gel results suggest an output factor 3% lower than the value recommended by the vendor (0.841 +/- 0.009 versus 0.870, respectively). CONCLUSIONS: A comparison with corresponding measurements published in the literature indicates that output factor results of this work are in agreement with those obtained using dosimetric systems which, besides fine spatial resolution and lack of angular and dose rate dependence of the dosimeter's response, share with polymer gels the favorable characteristic of minimal radiation field perturbation.
机译:目的:采用水当量聚合物凝胶剂量计和磁共振成像来测量伽马刀模型C放射牢房中可用的两个最小治疗领域的输出因子,采用4和8 mm最终准直器头盔的那些。方法:制备和照射三种荧光酚醛甘油凝胶制剂,使得测量输出因子的场的单次射击,并在每个探测器中测量其输出系数18mm场的单次拍摄。强调对减少的不确定性的细化数据处理方法的开发和基准,这完全利用在剂量计中登记的3D剂量分布。结果:发现8毫米准直器头盔的输出系数的高分子凝胶结果与供应商建议的相应价值密切一致(分别为0.955 +/- 0.007,分别为0.956)。然而,对于4mm准直仪头盔,聚合物凝胶结果表明输出因子低于供应商推荐的值(分别为0.870的0.841 +/- 0.009)。结论:与文献中发表的相应测量的比较表明,除了精细的空间分辨率和缺乏对剂量计的反应的角度和剂量率依赖性之外,该工作的输出因子结果与使用剂量计的响应缺乏,与剂量计的反应缺乏相一致。凝胶最小辐射场扰动的有利特征。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号