Thrombectomy for acute internal carotid thrombosis: five thrombectomy devices compared.

机译:急性内颈血栓形成的血栓切除术:比较了五种血栓切除术装置。

获取原文
获取原文并翻译 | 示例
           

摘要

Objective. To test the safety and efficiency of commercially available thrombectomy catheters in clearing simulated internal carotid artery (ICA) thrombosis. Design. Comparative in vitro study. Materials and methods. A model of the ICA was filled with human thrombus, the 'circle of Willis' back pressure was set at either 10 or 30 cm of water. Five thrombectomy devices (Hydrolyser, Clot Buster, Acolysis System, AngioJet and Fogarty embolectomy catheter) were compared for (i) efficiency at removing thrombus, (ii) pressure changes at the tip of each device, and (iii) distal embolisation by flow cytometry. Results. Thrombus clearance was greatest with the AngioJet (median 95%, range 92-97%) and least with the Acolysis System (median 34%, range 12-50%). The Clot Buster and Hydrolyser were safest as they produced only negative tip pressures, the AngioJet and Balloon catheter produced positive and negative pressures risking distal embolisation. The Acolysis system produced no pressure change during use. Distal embolisation (of particles between 5 and 40 microm diameter) was greatest with the Fogarty balloon catheter at 10 cm water (P<0.05) and least with the Hydrolyser and Clot Buster. Conclusion. Balloon embolectomy for ICA thrombosis risks further embolic cerebral damage. The Hydrolyser and the Clot Buster show the greatest promise for ICA thrombectomy.
机译:目的。为了测试市售血栓切除导管在清除模拟颈内动脉(ICA)血栓形成中的安全性和效率。设计。体外比较研究。材料和方法。 ICA模型充满了人的血栓,“威利斯循环”的背压设定为10或30厘米水柱。比较了五个血栓切除术设备(脱水机,Clot Buster,Acolysis系统,AngioJet和Fogarty栓塞术导管)的(i)清除血栓的效率,(ii)每个设备尖端的压力变化和(iii)通过流式细胞术进行远端栓塞。结果。 AngioJet的血栓清除率最高(中位数为95%,范围为92-97%),而Acolysis系统的血栓清除率最低(中位数为34%,范围为12-50%)。 Clot Buster和Hydrolyser最为安全,因为它们仅产生负压,而AngioJet和Balloon导管则产生正压和负压,有远端栓塞的危险。使用过程中,Acolysis系统不会产生压力变化。远端栓塞(直径在5至40微米之间的颗粒)在10 cm水位的Fogarty球囊导管最大(P <0.05),而在Hydrolyser和Clot Buster栓塞最少。结论。 ICA血栓球囊摘除术有进一步栓塞性脑损伤的风险。 Hydrolyser和Clot Buster显示了ICA血栓切除术的最大前景。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号