Electron ionization in LC-MS: recent developments and applications of the direct-EI LC-MS interface

机译:LC-MS中的电子电离:Direct-EI LC-MS接口的最新发展和应用

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this article is to underline the possibility of efficiently using electron ionization (EI) in liquid chromatography (LC) and mass spectrometry (MS). From a historical perspective, EI accompanied the first attempts in LC-MS but, owing to several technical shortcomings, it was soon outshined by soft, atmospheric pressure ionization (API) techniques. Nowadays, two modern approaches, supersonic molecular beam LC-MS and direct-EI LC-MS, offer a valid alterative to API, and preserve the advantages of EI also in LC-MS applications. These advantages can be summarized in three crucial aspects: automated library identification; identification of unknown compounds, owing to EI extensive fragment information; inertness to coeluted matrix interferences owing to very unlikely ion–ion and ion–molecule interactions in the EI gas-phase environment. The direct-EI LC-MS interface is a simple and efficient solution able to produce high-quality, interpretable EI spectra from a wide range of low molecular weight molecules of different polarity. Because of the low operative flow rates, this interface relies on a nano-LC technology that helps in reducing the impact of the mobile phase on the gas-phase environment of EI. This review provides an extensive discussion on the role of EI in LC-MS interfacing, and presents in detail several performance aspects of the direct-EI LC-MS interface, especially in terms of response, mass-spectral quality, and matrix effects. In addition, several key applications are also reported. Figure?Range of HPLC amenable compound classes by LC-EIMS
机译:本文的目的是强调在液相色谱(LC)和质谱(MS)中有效使用电子电离(EI)的可能性。从历史的角度来看,EI伴随着LC-MS的首次尝试,但由于一些技术缺陷,它很快被软性,大气压电离(API)技术所超越。如今,超声分子束LC-MS和直接EI LC-MS这两种现代方法提供了API的有效替代方案,并且在LC-MS应用中也保留了EI的优势。这些优势可以归纳为三个关键方面:图书馆自动识别;借助EI广泛的片段信息鉴定未知化合物;由于在EI气相环境中极不可能发生的离子-离子和离子-分子相互作用,因此对共洗脱基质干扰具有惰性。直接EI LC-MS接口是一种简单有效的解决方案,能够从各种极性不同的低分子量分子中产生高质量,可解释的EI光谱。由于操作流速低,该界面依赖于纳米液相色谱技术,该技术可帮助减少流动相对EI气相环境的影响。这篇综述对EI在LC-MS接口中的作用进行了广泛的讨论,并详细介绍了直接EI LC-MS接口的几个性能方面,特别是在响应,质谱质量和基质效应方面。此外,还报告了一些关键应用。图LC-EIMS对HPLC适用化合物类别的范围

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号