Towards compound-independent calibration for organic compounds using online isotope dilution mass spectrometry

机译:使用在线同位素稀释质谱法实现有机化合物的化合物独立校准

获取原文
获取原文并翻译 | 示例
           

摘要

Isotope dilution mass spectrometry (IDMS) can be considered a primary measurement method directly traceable to the International System of Units (SI). This measurement technique is increasingly employed in routine laboratories, owing to its unequalled analytical performance, precision and ease of accreditation. Unfortunately, for the adequate application of IDMS, several isotopically labelled standards, corresponding to the compounds of interest, are required. Additionally, when the enriched isotope is continuously added after a chromatographic separation, and an elemental ion source is used, it allows quantification of the different analytes being eluted from the column without requiring specific standards for each compound (online IDMS). In this article, we discuss how the traditional applicability of online IDMS for elemental speciation can be dramatically expanded by using carbon isotope tracers, oxidation or combustion reactions and a conventional molecular ion source. With such a strategy every carbon-containing compound being eluted from a chromatography system can be quantified without the need for specific standards as long as quantitative combustion/oxidation and complete elution occur. So far, only gas chromatography-combustion-mass spectrometry applications have been described, but recent results indicate the great possibilities of extending this novel approach to the quantification of organic compounds after separation by liquid chromatography.
机译:同位素稀释质谱法(IDMS)可被视为直接溯源到国际单位制(SI)的主要测量方法。由于其无与伦比的分析性能,精确度和易于鉴定,该测量技术越来越多地用于常规实验室。不幸的是,为了适当地使用IDMS,需要几种与目标化合物相对应的同位素标记标准品。此外,在色谱分离后连续添加富集的同位素并使用元素离子源时,它可以对从色谱柱洗脱的不同分析物进行定量,而无需为每种化合物指定特定的标准(在线IDMS)。在本文中,我们讨论了如何通过使用碳同位素示踪剂,氧化或燃烧反应以及常规的分子离子源来显着扩展在线IDMS对元素形态的传统适用性。通过这种策略,只要发生定量燃烧/氧化和完全洗脱,就可以对从色谱系统洗脱的每种含碳化合物进行定量,而无需特定标准。到目前为止,仅描述了气相色谱-燃烧-质谱法的应用,但是最近的结果表明将这种新方法扩展到液相色谱分离后有机化合物定量的巨大可能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号