A novel absorption detection method for highly multiplexed capillary electrophoresis is presented for zone electrophoresis and for micellar electrokinetic chromatography. The approach involves the use of a linear photodiode array on which a capillary array is imaged by a camera lens. Either a tungsten lamp or a mercury lamp can be used as the light source such that all common wavelengths for absorption detection are accessible by simply interchanging narrow-band filters. Each capillary spans several diodes in the photodiode array for absorption measurements. Over 100 densely packed capillaries can be monitored by a single photodiode array element with 1024 diodes. The detection limit for rhodamine 6G for each capillary in the multiplexed array is similar to 1.8 x 10(-8) M injected (S/N = 2). The cross-talk between adjacent capillaries is less than 0.2%. Simultaneous analysis of 96 samples is demonstrated. [References: 20]
机译:提出了一种新的用于高度复用毛细管电泳的吸收检测方法,用于区域电泳和胶束电动色谱。该方法涉及使用线性光电二极管阵列,在该线性光电二极管阵列上通过照相机镜头将毛细管阵列成像。钨灯或水银灯都可以用作光源,这样,通过简单地互换窄带滤光片就可以访问所有用于吸收检测的通用波长。每个毛细管跨越光电二极管阵列中的几个二极管以进行吸收测量。一个带有1024个二极管的光电二极管阵列元件可以监视100多个密集包装的毛细管。在多路复用阵列中,每个毛细管的若丹明6G的检出限类似于注入的1.8 x 10(-8)M(S / N = 2)。相邻毛细管之间的串扰小于0.2%。同时分析了96个样品。 [参考:20]