首页> 外文期刊>Analytical chemistry >Ultrafast Multiplexed-Allergen Detection through Advanced Fluidic Design and Monolithic Interferometric Silicon Chips
【24h】

Ultrafast Multiplexed-Allergen Detection through Advanced Fluidic Design and Monolithic Interferometric Silicon Chips

机译:超快复用 - 过敏原通过先进的流体设计和单片干涉硅芯片检测

获取原文
获取原文并翻译 | 示例
       

摘要

A silicon-based miniaturized sensor chip combined with an advanced microfluidic module for the simultaneous, label-free immunochemical determination of four allergens, bovine milk protein, peanut protein, soy protein, and gliadin, is presented. The sensor chip consists of an array of 10 broad-band Mach-Zehnder interferometers (BB-MZIs) monolithically integrated on silicon, along with their respective broad-band light sources. The BB-MZIs were biofunctionalized with the targeted allergens and their responses during immunoreaction were monitored by multiplexing their transmission spectra through an external miniaturized spectrometer. The assay is performed by running mixtures of calibrators or samples with the antibodies against the four allergens followed by an antispecies specific antibodies solution. Employing a fluidic module of nearly one-dimensional geometry, that provided for uniform delivery of the reagents, CV values 6% were achieved for the responses of the 10 BB-MZIs, allowing for reliable multianalyte determinations. The analysis is completed in 6.5 min, and the detection limits were 0.04 mu g/mL for bovine k-casein, 1.0 mu g/mL for peanut protein, 0.80 mu g/mL for soy protein, and 0.10 mu g/mL for gliadin. The assays were accurate (recoveries 88-118%) and repeatable (intra- and interassay CVs 7% for all four allergens). Finally, the sensor was evaluated by analyzing samples from a cleaning in place system (CIP) of a dairy industry and the results obtained were in good agreement with those received by the respective ELISAs. The analytical characteristics of the sensor combined with the short analysis time and the small chip size make the proposed system an ideal tool for on-site multianalyte determinations.
机译:提出了一种基于硅基小型化传感器芯片,用于同时,无标度免疫化学测定的四种过敏原,牛奶蛋白,花生蛋白,大豆蛋白和胶质苷的先进的微流体模块。传感器芯片由一系列10个宽带马赫 - Zehnder干涉仪(BB-MZIS)的阵列组成,其在硅片上整合在一起,以及它们各自的宽带光源。 BB-MZIS通过靶向过敏原生物官能化,并通过外部小型化光谱仪复用其透射光谱来监测其在免疫反应期间的反应。通过将校准剂或样品与抗体的抗体与四个过敏原的抗体一起运行,然后进行抗分析特异性抗体溶液来进行测定。采用具有近一维几何形状的流体模块,其用于均匀递送试剂,对于10bb-mzis的响应,实现了CV值的CV值。允许可靠的多分析测定。分析在6.5分钟内完成,牛k-casein的检测限为0.04μg/ ml,花生蛋白为1.0μmg/ ml,对于大豆蛋白0.80μg/ ml,胶合素0.10μg/ ml 。测定是准确的(回收88-118%)和可重复的(所有四个过敏原的和干扰和间隙和间隙7%)。最后,通过分析来自乳制品行业的清洁系统(CIP)的清洁系统(CIP)的样品来评估传感器,并且获得的结果与各自的ELISA收到的结果良好。传感器的分析特性与短分析时间和小芯片尺寸结合,使提出的系统成为现场多分析测定的理想工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号