Dried blood spots (DBS) are often used as a less invasive alternative to venous blood sampling. Despite its numerous advantages, the use of conventional DBS suffers from the hematocrit (hct) effect when analyzing a subpunch. This effect could be avoided by using hct-independent sampling devices, of which the hemaPEN is a recent example. This device collects the blood via four integrated 2.74 mu L microcapillaries, each depositing the blood on a prepunched paper disc. In this study, we evaluated the technical performance of the hemaPEN devices, using an extensive bioanalytical validation and application on authentic patient samples. An LC-MS/MS method quantifying caffeine and its metabolite paraxanthine in dried whole blood (using the hemaPEN device) was fully validated, meeting all preset acceptance criteria. A comparative analysis of 91 authentic patient samples (hct range: 0.17-0.53) of hemaPEN, 3 mm DBS subpunches, and whole blood revealed a limited hct dependence (<= 7% concentration difference over a 0.20-0.50 hct range) for the hemaPEN devices, which we could not attribute to the analytical procedure. Using conventional partial-punch DBS (3 mm punches), concentration differences of >= 25% over this hct range were found. The hemaPEN showed to be robust to the effects of blood sample volume, device lot, analytical operator, and storage stability. The technical performance of the hemaPEN when dealing with patients having a high hct and in cases where a large blood drop is present should be further investigated. Based on the successful validation and application on patient samples, we conclude that the hemaPEN device shows good potential for the volumetric collection of DBS.
机译:干血斑(DBS)通常用作静脉血液采样的较少侵入性替代品。尽管有许多优点,但在分析子统计时,常规DB的使用患有血细胞比容(HCT)效应。通过使用HCT独立的采样装置可以避免这种效果,其中血红是最近的一个例子。该装置通过四个集成的2.74 mu l微孔收集血液,每个微毛细管每次沉积在纸张上的血液上。在本研究中,我们评估了血红素设备的技术性能,使用广泛的生物分析验证和在真实的患者样品上应用。 LC-MS / MS方法量化咖啡因及其在干燥的全血(使用血红素装置)中的代谢物帕拉蒂嗪被完全验证,满足所有预设的验收标准。 91正宗患者样品(HCT范围:0.17-0.53)的比较分析血红素,3mM DBS沉积和全血显示出血红素的有限的HCT依赖性(<= 7%浓度差异为0.20-0.50小时范围)设备,我们无法归因于分析程序。使用常规的部分冲头DB(3mm打孔),发现在该HCT范围内> = 25%的浓度差异。血红表现为血液样品体积,装置批,分析算子和储存稳定性的影响是强大的。应进一步研究处理具有高HCT的患者时血红素的技术性能。基于患者样品的成功验证和应用,我们得出结论,血红素装置显示出DB的体积收集的良好潜力。

著录项

  • 来源
    《Analytical chemistry》 |2019年第22期|共9页
  • 作者单位

    Univ Ghent Fac Pharmaceut Sci Dept Bioanal Lab Toxicol Ottergemsesteenweg 460 B-9000 Ghent Belgium;

    Univ Santiago de Compostela Fac Med Inst Forens Sci Toxicol Serv Rua San Francisco S-N Santiago De Compostela 15782 Spain;

    Univ Ghent Fac Pharmaceut Sci Dept Bioanal Lab Toxicol Ottergemsesteenweg 460 B-9000 Ghent Belgium;

    Univ Ghent Fac Pharmaceut Sci Dept Bioanal Lab Toxicol Ottergemsesteenweg 460 B-9000 Ghent Belgium;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 分析化学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号