首页> 外文期刊>Journal of computer assisted tomography >Low-dose chest computed tomography with sinogram-affirmed iterative reconstruction, iterative reconstruction in image space, and filtered back projection: studies on image quality.
【24h】

Low-dose chest computed tomography with sinogram-affirmed iterative reconstruction, iterative reconstruction in image space, and filtered back projection: studies on image quality.

机译:低剂量胸部计算断层扫描与铭文肯定的迭代重建,图像空间中的迭代重建,并过滤后投影:图像质量研究。

获取原文
获取原文并翻译 | 示例
           

摘要

This study aimed to determine optimal strength of sinogram-affirmed iterative reconstruction (SAFIRE) and to evaluate image quality (IQ) of low-dose chest computed tomography (LDCT) using SAFIRE compared with iterative reconstruction in image space (IRIS) and filtered back projection (FBP).Thirty patients underwent LDCT. Computed tomography (CT) was reconstructed using 5 strengths of SAFIRE (S1-S5), IRIS, and FBP. Objective noise of CT was measured. Two radiologists evaluated CT for subjective IQ, beam-hardening artifacts, and overall IQ.Measured noise was highest in FBP, followed by S1, S2, S3, IRIS, S4, and S5. S2 and S3 demonstrated significantly higher overall IQ scores than the other strengths (P < 0.05). Overall IQ and beam-hardening artifacts of S2 and S3 were significantly better than those in FBP and IRIS (P < 0.001).S2 or S3 strengths of SAFIRE can be used practically in clinical routines and may have more potential than IRIS and FBP for LDCT with improved IQ.
机译:本研究旨在确定铭记肯定的迭代重建(SAFIRE)的最佳强度,并使用Safire评估低剂量胸部计算机断层扫描(LDCT)的图像质量(IQ)与图像空间(虹膜)中的迭代重建和滤波后投影相比 (FBP)。患者接受了LDCT。 使用5个Safire(S1-S5),IRIS和FBP重建计算断层扫描(CT)。 测量CT的客观噪声。 两个放射科医生用于主观IQ,光束硬化伪像和总体IQ的CT,在FBP中最高,其次是S1,S2,S3,虹膜,S4和S5。 S2和S3显示出比其他强度的总IQ分数明显更高(P <0.05)。 S2和S3的总体IQ和光束硬化伪像明显优于FBP和IRIS(P <0.001).s2或S3的安全性,可以在临床常规中实际使用,并且可能比虹膜和LDCT的FBP更多潜力 改善智商。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号