首页> 外文期刊>Proceedings of the Institution of Mechanical Engineers, Part H. Journal of Engineering in Medicine >Effect of filling pressure in jetting dispenser on the performance of blood glucose test strips using immersion gold-plated printed circuit board
【24h】

Effect of filling pressure in jetting dispenser on the performance of blood glucose test strips using immersion gold-plated printed circuit board

机译:采用浸渍镀镀镀金印刷电路板灌装压力灌装压力对血糖试验条性能的影响

获取原文
获取原文并翻译 | 示例
           

摘要

To ensure accurate glucose readings when dispensing glucose oxidase enzyme solution from a jetting dispenser onto glucose test strips fabricated from an immersion gold-plated printed circuit board, every drop of the enzyme solution needs to have nearly the same weight and to be dispensed on the reaction zone of the test strips. Experimental results in this study show that the filling pressure in the fluid reservoir containing the glucose enzyme solution to dispense onto the test strips significantly affect the glucose test results. A filling pressure of 12?psi produces test strips with lower coefficient of variation and standard deviation than 10 and 14?psi. Proper filling pressure for dispensing glucose enzyme onto glucose test strips needs to be determined for any enzyme compound formulation.
机译:为了确保从喷射分配器将葡萄糖氧化酶酶溶液分配到由浸泡镀金印刷电路板制造的葡萄糖测试条上的葡萄糖试验条上时,每下需要具有几乎相同的重量并在反应上分配时,以确保精确的葡萄糖读数。 测试条带。 该研究的实验结果表明,含有葡萄糖酶溶液的流体贮存器中的填充压力显着影响葡萄糖测试结果。 填充压力为12?psi产生具有较低变异系数和标准偏差的测试条,而标准偏差小于10和14.PSI。 需要针对任何酶化合物制剂测定用于将葡萄糖酶分配到葡萄糖试验条上的适当填充压力。