Self-measurement with Icare HOME tonometer, patients’ feasibility and acceptability

机译:用ic ic ob tonometer,患者的可行性和可接受性自测

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To evaluate and compare the accuracy of self-measurement of intraocular pressure using Icare Home rebound tonometer with Goldmann applanation tonometer and assess acceptability of self-tonometry in patients with glaucoma and ocular hypertension. Methods: In the study, 117 subjects were trained to use Icare Home for self-measurement. Icare Home tonometer readings were compared with Goldmann applanation tonometer, including one eye per patient. Agreement between the two methods of measurement was evaluated by Bland and Altmann analysis. Questionnaire was used to evaluate patients’ perception of self-tonometry. Results: One hundred and three out of 117 patients (88%) were able to measure their own intraocular pressure and 96 (82%) fulfilled the requirements for certification. The mean (SD) difference Goldmann applanation tonometer minus Icare Home was 1.2 (2.4)?mmHg (95% limits of agreement, –3.4 to 5.9?mmHg). The magnitude of bias between the two methods depended on central corneal thickness, with greater bias at central corneal thickness <500?μm. In 65 out of 96 subjects (67.7%), Icare Home results were within 2?mmHg of the Goldmann applanation tonometer. Seventy-three out of 93 (78.5%) felt that self-tonometry was easy to use and 75 patients (80.6%) responded that they would use the device at home. Conclusion: Icare Home tonometry tends to slightly underestimate intraocular pressure compared to Goldmann applanation tonometer. Most patients were able to perform self-tonometry and found it acceptable for home use. Measurements using rebound self-tonometry could improve the quality of intraocular pressure data and optimize treatment regimen.
机译:目的:使用ICare Home Rebound Tonometer评估和比较眼压的自测量的准确性,并使用Goldmann Applanation Tonometer评估血糖瘤和眼高压患者自我读数的可接受性。方法:在研究中,培训117名科目以使用ICare Home进行自我测量。与Goldmann Applanation Tonometer相比,ICare Home Tonometer读数,包括每位患者的一只眼睛。通过平淡和Altmann分析评估两种测量方法之间的协议。调查问卷用于评估患者对自我读数的看法。结果:117名患者中的一百三分之一(88%)能够测量其自己的眼压压力,96(82%)满足认证要求。平均值(SD)差异Goldmann Applanation Tonometer减去ICare Home为1.2(2.4)?MMHG(95%的协议限制,-3.4至5.9?mmhg)。两种方法之间的偏压尺寸依赖于中心角膜厚度,在中央角膜厚度<500Ωμm处具有更大的偏压。在96名受试者中有65名(67.7%),ICare Home业绩在2?Mmhg的Goldmann Applanation Tonometer之内。 93中的七十三(78.5%)觉得自我读数易于使用,75名患者(80.6%)回应,他们将在家里使用该装置。结论:与Goldmann Applanation Tonometer相比,ICare Home Tononetry往往略微低估了眼压。大多数患者能够进行自我曲目,发现它可以接受家庭使用。使用反弹自我测量测量可以提高眼内压力数据的质量,优化治疗方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号